What to do in the event of malfunctions
S100 / OPMI pico
Version 9.0
Page 214
G-30-1781-en
Streaming does not work.
1920 x 1080 resolution too
high for video.
Switch HDTV format to
"720p" (Settings menu >
System submenu > Video).
Video quality too high.
Set video quality to "low"
(Settings menu > Record
submenu).
Network settings in receiver
computer wrong.
Check network settings
(Settings menu > System
submenu > Network).
Firewall prevents data flow.
Check firewall settings.
–
Receiver software does not
support RTSP / RTP or is
configured incorrectly.
Check network settings
(Settings menu > System
submenu > Network).
Check receiver software
settings.
–
Fault
Possible reason
Corrective measure
Cross ref-
erence
Summary of Contents for S100
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Page 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Page 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Page 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Page 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Page 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Page 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Page 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Page 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Page 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...