Operation
S100 / OPMI pico
Version 9.0
Page 196
G-30-1781-en
The following settings are preinstalled settings. These settings can be
changed by the IT Administrator any time that the network environment
requires it.
1
IP settings (static)
IP address: 192.168.0.11
Subnet mask: 255.255.255.0
Standard gateway: 192.168.0.1
2
Network share
Host IP: 192.168.0.10 (external computer)
Share name: recording
Work group: (or domain name)
User name: zeiss
Password: recording
3
Streaming
Client IP: 192.168.0.255 (broadcast- or individual client IP address)
Port: 5004 (does not normally need to be changed)
You can stream to a single receiver (client IP) by indicating a single IP
address. But the stream can also be transmitted to several receivers at the
same time if you use the broadcast address of your network area as the
client IP. In addition, each receiver must set up the connection via RTSP.
Use broadcast only in a cable-connected network and not to WLAN
receivers.
If streaming is active in the S100 / OPMI pico, the following URL can be
used in the video software (additional software may be required):
URL: rtsp://192.168.0.11:8554/video
You only have to adjust the IP address or replace the name configured
in the Domain Name System (DNS).
You can read out the IP address of the surgical microscope using the
"IP Address" button even if DHCP is active (see Page 193).
4
Web control / Web interface
URL: http://192.168.0.11
(Port 80 must not be displayed)
You can read out the IP address of the surgical microscope using the
"IP Address" button even if DHCP is active (see Page 193).
If a name has been saved in your Domain Name System (DNS), you can
also use this name instead of the IP address.
If you have any questions, contact your IT Administrator.
Summary of Contents for S100
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Page 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Page 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Page 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Page 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Page 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Page 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Page 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Page 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Page 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...