What to do in the event of malfunctions
S100 / OPMI pico
Version 9.0
Page 216
G-30-1781-en
Lamp failure of halogen light source
The lamp housing contains a backup lamp which can be manually swung
into the illumination beam path when the first lamp fails. The open flap (1)
indicates that the backup lamp is in operation.
•
Press button (2) to manually swivel the backup lamp into the beam path.
The open flap (1) indicates that the backup lamp is in operation.
•
Replace the defective lamp at the first opportunity (see Page 224).
•
If the backup lamp is defective and replacement is not possible, continue
surgery using an external OR illuminator.
Fig. 99: Swiveling in the
backup lamp
CAUTION
Risk of burns!
In the event of overheating, the lamp switches off automaticall.
When the lamp has cooled down, it switches on again automatically.
If the microscope is positioned unattended above the patient, the patient
can be burnt due to the heat generated.
•
Immediately activate the backup lamp after a lamp failure or switch
off the light source completely.
•
Check if the light source is switched off before leaving the device
unattended.
1
2
Summary of Contents for S100
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 9 0 2018 11 23...
Page 4: ...S100 OPMI pico Version 9 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMI pico Version 9 0 Page 6 G 30 1781 en...
Page 49: ...Version 9 0 G 30 1781 en Page 49 S100 OPMI pico Safety measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety measures S100 OPMI pico Version 9 0 Page 50 G 30 1781 en...
Page 88: ...Device overview S100 OPMI pico Version 9 0 Page 88 G 30 1781 en...
Page 115: ...Version 9 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 9 0 Page 126 G 30 1781 en...
Page 168: ...Operation S100 OPMI pico Version 9 0 Page 168 G 30 1781 en...
Page 220: ...What to do in the event of malfunctions S100 OPMI pico Version 9 0 Page 220 G 30 1781 en...
Page 264: ...Indexes S100 OPMI pico Version 9 0 Page 264 G 30 1781 en...
Page 276: ...Keyword index S100 OPMI pico Version 9 0 Page 276 G 30 1781 en...
Page 277: ...Version 9 0 G 30 1781 en Page 275 S100 OPMI pico Blank page for your notes...