ZEISS
FIRST-TIME SET-UP
Axio Imager 2
VIS-LED attachment lamp for transmitted light
64
430000-7544-001
01/2016
3.31
VIS-LED attachment lamp for
transmitted light
3.31.1
The VIS-LED attachment lamp is
only intended for use with
transmitted light.
The VIS-LED attachment lamp is only intended for
transmitted light operation.
Do not look directly into the LED light.
Only insert or remove the VIS-LED
attachment lamp in / from the
transmitted-light port while switched
off.
•
Loosen clamping screw (Fig. 50/
1
) on
transmitted-light port (Fig. 50/
2
). Remove the
halogen lamp.
•
Use an AF 3 ball-headed screwdriver and
clamping screw (Fig. 50/
1
) to lock the
attachment lamp (Fig. 50/
3
) with dovetail
mount firmly into socket (Fig. 50/
2
).
•
Connect the three-pin illuminator plug
(Fig. 50/
5
) to the lower three-pole 12 V / 100 W
socket (Fig. 50/
6
) for transmitted light (
TL
) on
the back of the instrument.
•
For the coded stand, set the transmitted-
light/reflected-light toggle switch to the position
for transmitted light (
TL
). If the motorized stand
is used, switch between reflected light and
transmitted light using the TFT display touch
screen.
If the AxioCam MRm is used for documentation purposes, the 12 % gray filter provided must
be employed to attenuate light intensity.
The add-on illuminator is equipped with a BNC port (Fig. 43/
4
) for TTL trigger control.
Fig. 50
Installing the VISLED attachment
lamp