OPERATION
Axio Imager 2
Operating the motorized microscope via the touchscreen of the TFT display
130
430000-7544-001
01/2016
Only dyes may be selected instead of fluorescence filter sets. In this case, the user chooses the relevant
dye from the subselection options listed.
Fluorescence filter sets can be selected by their numbers. Additionally, a suitable dye may be picked for
each filter set from the submenu.
Once a new reflector has been allocated, the relevant reflector button on the
Microscope
operation
page will show the corresponding information.
Further information about the programming of rewritable ACR reflector modules can be found
Configuring a reflector position, using Colibri
as a source of illumination
Once Colibri has been recognized and configured
as a source of illumination, an appropriate LED can
be selected for each filter set or dye.
•
Press the
Colibri
button.
•
In the following pop-up menu (Fig. 135) select
one or more LEDs to match the selected
reflector position.
Alternatively, an optional white-light source
(HXP 120) can also be selected. Note that LED
and HXP cannot be chosen at the same time.
•
Acknowledge the selection procedure by
pressing
Save
or press
Cancel
to discard the
selection.
•
Acknowledge the selection procedure a second
time by pressing
Save
or press
Cancel
to
discard the selection.
•
Acknowledge the confirmation prompt with
OK
.
Fig. 134
Configuring the reflector position
using Colibri
Fig. 135
Selecting Colibri / HXP 120
illumination