Axio Imager 2
OPERATION
ZEISS
Operating the motorized microscope via the touchscreen of the TFT display
01/2016
430000-7544-001
111
−
Make It Visible!
Button (Fig. 100/
3
)
This button switches the microscope to a basic state:
Transmitted-light lamp set to medium intensity (2.3 V),
Luminous-field diaphragm opened,
Aperture diaphragm opened
TL shutter open,
RL shutter closed,
All filter wheels in transmitted light switched to blank aperture (100% directed to eyepieces),
Condenser switched to brightfield,
Reflector turret switched to the nearest HAL- position (Halogen = transmitted light),
Light path switched 100% to eyepieces.
4.8.4
Favorite page
In order to increase the operation efficiency of the
microscope, a frequently used page can be set as a
favorites page. When selected, the system jumps
directly to this page. For the procedure of how to
define this page, see also Section 4.8.6.
4.8.5
Microscope page
By pressing the
Microscope
button in the
navigation bar on the
Home
page, the user can
access the microscope operation pages (compare
Fig. 103):
−
Turret
−
Illumination
−
Automatic operation
−
XYZ
Fig. 101
Favorites page
Fig. 102
Selecting the Microscope page