Axio Imager 2
FIRST-TIME SET-UP
ZEISS
Mechanical stage 75x50 mot. CAN
01/2016
430000-7544-001
77
3.37
Mechanical stage 75x50 mot. CAN
To position the 75x50 mot. CAN stage,
−
the coaxial drive,
−
the trackball or
−
the joystick
can be used.
3.37.1
Assembling and connecting 75x50 mot. CAN mechanical stage with coaxial drive
On delivery, the mechanical stage has a transport lock. This should be removed before attaching
the stage, as otherwise the mechanical stage could be damaged.
•
Unpack the mechanical stage and remove the transport lock (Fig. 70/
2
).
•
Slide the stage plate (Fig. 70/
1
) forwards and fasten coaxial drive (Fig. 70/
4
) to the underside of the
stage using two screws (Fig. 70/
5
).
•
Place mechanical stage (Fig. 70/
1
) onto the stage carrier so that the holes in the underside of the
mechanical stage coincide with the through holes of the stage carrier.
•
Using the angled Allen key (AF 3), screw four screws (Fig. 70/
6
) from the bottom into the underside of
the stage, with the shorter screws being inserted in the front holes.
1
Mechanical stage 75x50 mot. CAN
2
Transport lock (screw)
3
Angled CAN bus connector of the connection cable (100-0600-144)
4
Electronic coaxial drive CAN
5
Fastening screws on coaxial drive
6
Fastening screws on stage carrier
Fig. 70