FIRST-TIME SET-UP
Axio Imager 2
Screwing in objectives
40
430000-7544-001
01/2016
3.9
Screwing in objectives
•
Move the mechanical stage with stage carrier to
the lower stop position.
•
Remove dust covers (Fig. 17/
2
) from the
respective openings on the objective nosepiece.
•
Remove objectives (Fig. 17/
1
) from the case and
screw them into the nosepiece (Fig. 17/
3
)
starting with the lowest magnification
(clockwise rotation).
•
Make sure that the unused objective mounts of
the nosepiece are covered by dust caps.
Use adapter M27x0.75 "0" on W 0.8
(000000-1095-168) in combination
with objectives with W 0.8" thread.
3.10
Inserting reflector turret, 6x20
compensator mount or 4-position
modulator turret
Install or change the motorized
reflector turret or modulator turret
only when the microscope is switched
off.
•
Loosen the clamping screw (Fig. 18/
1
). Pull out
the reflector turret (Fig. 18/
6
) or the cover part,
using the grip recess (condition on delivery).
•
Insert the compensator mount (Fig. 18/
5
) or the
4-position modulator turret (Fig. 18/
7
) manual
or motorized) fully into guide (Fig. 18/
4
) on the
top part of the stand. Tighten clamping screw
(Fig. 18/
2
).
•
Attach reflector turret (Fig. 18/
6
) to the guide
3
) in the top part of the stand and push
it in until it (audibly) snaps in.
•
Tighten clamping screw (Fig. 18/
1
).
•
If the reflector turret was supplied without
inserted reflector modules, insert them as
described in Section 3.11.
If the instrument is to be transported to another location, the reflector turret must be replaced
again by the cover part with grip recess.
Fig. 17
Screwing in objectives
Fig. 18
Inserting reflector turret,
compensator mount or modulator
turret