ZEISS
FIRST-TIME SET-UP
Axio Imager 2
Attaching the condenser carrier to the transmitted-light / reflected-light stage carrier
50
430000-7544-001
01/2016
3.20
Attaching the condenser carrier to
the transmitted-light / reflected-
light stage carrier
•
Use the drive knob (Fig. 32/
2
) to adjust the
guide of the condenser carrier (Fig. 32/
3
) in
such a way that the two screws (Fig. 32/
4
) are
accessible.
•
Attach the condenser carrier to the stage carrier
(Fig. 32/
1
). Push it firmly in a straight line to the
upper stop, and tighten the two screws.
3.21
Changing the fine-drive knob or
the fine-drive disk on the stand
The fine-focusing drive to the right and left on the
stand can be provided either with a fine-drive knob
or a fine-drive disk.
To change the fine-drive knob or the fine-drive
disk, proceed as follows:
•
Turn the coarse-drive knob on the relevant side
of the stand so that the mounting hole
(Fig. 33/
2
) points upward.
•
Insert the ball-headed screwdriver SW 1.5
(Fig. 33/
1
) into the mounting hole and loosen
the locking screw of the fine-drive knob or the
fine-drive disk.
•
Pull the fine-drive knob or the fine-drive disk
sideways off the shaft.
•
Slide the fine-drive disk or the fine-drive knob
onto the shaft, so that the recess (Fig. 33/
3
or
4
) is located under the mounting hole.
•
Tighten the locking screw.