Reichert Manual Lensometer ML1 User Manual Download Page 1

ML1

Manual Lensometer

®

 

User’s Guide

Summary of Contents for Manual Lensometer ML1

Page 1: ...ML1 Manual Lensometer User s Guide...

Page 2: ...to make changes in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 9001 13485 Certified Reichert products are designed and...

Page 3: ...12 Lens Analysis 12 Reading in Plus Cylinder 13 Reading in Minus Cylinder 13 Lens Centering Marking 13 Measuring Prism Power 14 Measuring Bifocal Additions 15 Fused Bifocal Lens 15 One Piece Bifocal...

Page 4: ...STATIC DISCHARGE SENSITIVE DEVICES ESDS THAT MAY BE SENSITIVE TO STATIC CHARGES PRODUCED BY THE HUMAN BODY DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER PRECAUTIONS Warning THE EQUIPMENT OR SYSTEM SH...

Page 5: ...BY THE HUMAN BODY DO NOT REMOVE THE COVERS WITHOUT TAKING PROPER ESDS PRECAUTIONS Caution do not use solvents or strong cleaning solutions on any part of this instrument AS damage to the unit may occ...

Page 6: ...t are included in the User s Guide Type B Product Classification Class I Equipment Continuous Operation Protective Earth ON OFF DC Volts Date of Manufacture REF Catalog Number SN Serial Number Waste o...

Page 7: ...g and reference manual for operation maintenance and troubleshooting We recommend that you read it carefully prior to use and follow the instructions in the guide to ensure optimum performance of your...

Page 8: ...it 5 upright on a flat surface Remove the accessories Refer to Figure SU 03 6 Accessories Included 2 AAA Batteries P N 13950000 902 Dust Cover P N 15110 016 Contact Lens Holder P N 15110 017 User s Gu...

Page 9: ...Eyepiece 5 Prism Compensator 6 Lens Clamp 7 Lens Holder 8 Power Wheel 9 Axis Wheel 10 Battery Compartment 11 Tilt Lock Lever 12 Lens Table 13 Figure SU 05 ML1 Parts Diagram 12 7 9 11 5 10 8 6 7 4 3 2...

Page 10: ...o loosen the tilt mechanism Refer to Figure SU 08 Move the main body up or down to the desired 2 position Tighten the Tilt Lock Lever to secure the main 3 body in the desired position Refer to Figure...

Page 11: ...Slowly turn the Eyepiece clockwise until the 4 black reticle lines are in sharp focus Refer to Figure SU10 Note Do not go beyond that point Remove the paper 5 Press the ON OFF Switch The green LED 6 l...

Page 12: ...Note Make certain that the last focusing movement of the Power Wheel is topside toward you If both the Sphere Lines 3 narrow lines and 2 Cylinder Lines 3 wide lines come into focus at the same time t...

Page 13: ...nses Example of reading Plus Cylinder First reading 3 00D X 110 Second Reading 2 50D Direction Rx 3 00 5 X 110 Example of reading in Minus Cylinder First reading 2 50D X 20 Second reading 3 00D Direct...

Page 14: ...le if the center of the target is above the center of the reticle base up prism is indicated Rotate the Prism Scale until the line of the scale 4 passes through the center of the target Refer to Figur...

Page 15: ...rtion of the lens Measure the power through the reading portion 3 Note The add power is the difference between these two readings Measuring Contact Lenses Remove the spectacle Lens Holder 1 Refer to F...

Page 16: ...y moistened with a commercial lens cleaner clean the lens surfaces Note Clean exterior surfaces only Note Do not remove any lenses Ink Pad Refilling Caution Do not overfill the inkwell It is only nece...

Page 17: ...e the Axis Wheel while observing the target If the center of the target remains stationary proceed with the target centering procedure Note If the center of the target moves with respect to the Eyepie...

Page 18: ...center of lens Prism Compensator not at 0 Set Prism Compensator to 0 Target not centered Center the target Marker out of alignment Have instrument serviced Target not centered with no lens in place ta...

Page 19: ...to 2 diopters 1 diopter beyond 2 diopters Cylindrical Axis Prism Compensator Range 1 to 180 degrees Diopter Range 15 prism diopters Step 1 degree Graduations 1 prism diopter steps Angle scale 0 degre...

Page 20: ...uct PATENT WARRANTY If notified promptly in writing of any action brought against the purchaser based on a claim that the instrument infringes a U S Patent Reichert will defend such action at its expe...

Page 21: ...21 15110 101 Rev D Notes...

Page 22: ...22 15110 101 Rev D Notes continued...

Page 23: ...23 15110 101 Rev D Notes continued...

Page 24: ...15110 101 Rev D January 21 2013...