SEM (JEOL, JSM-6301) Operation manual      2014 06 11 

   

The turning on the SEM 

1.

 

Turn on the scanning image observation instrument with the 

OPE

 

POWER

 switch at the rear of 

the instrument. 

2. Rotate the 

POWER

 key switch to the 

START

 position, and release the hand from the key. 

   

(The key rotates to the 

ON

 position automatically.) 

3. The 

ACCEL VOLTAGE READY

 lamp turns on at once. (The user should confirm the lamp 

turning on.) 

 

*Please contact the staff, if the lamp does not turn on. 

 

The sample exchange (The insert of the sample) 

1. Place the sample in the sample holder (cylinder), and align the sample upper face and the upper 

side of the cylinder, using the height adjusting screw (tighten by the screw). 

2. Screw the sample exchange rod to the sample holder, and tighten the rod.   
3. Put the sample stage to the position where the sample is possible to exchange.     

 

    Confirm 

the position for the sample exchange:

 

X:25.0  Y:35.0  TILT:000 

 

WD:39

 

4. Pull the exchange rod, and then fix the sample holder to the stopper. 
5. Fit the exchange rod to the sample exchange chamber. 

    While supporting the exchange rod by hand, press the vacuum operation button. 

6. When the lamp is turned off, rotate anticlockwise immediately the knob of the gate valve of the 

sample exchange chamber, and then pull out the knob fully.   

7. While looking the inner part of the chamber, push in the rod slowly, and fix the sample holder 

onto the stage. 

8. Rotate anticlockwise the rod, and remove the rod from the sample holder. 

        Pull out the sample rod from the chamber fully. 

9. Push in the knob of the gate valve of the sample exchange chamber, and rotate clockwise the 

knob. This operation makes a clicking sound.   

10. While supporting the exchange rod by hand, remove the rod by pressing the vacuum operation 

button. 

 

The observation with the SEM 

 

1. Press the 

F2 

function key on the keyboard. 

 

2. Confirm the accelerating voltage

 

being 

0.0 kV

 (see the second line from the top on the right 

side screen).   

          Turn on the

 ACCEL VOLTAGE

 switch. The extracting voltage is applied, and the emission 