Appendix
153
Care should be taken if the System is used adjacent to other elec-
trical equipment; if adjacent use is inevitable, such as in work
environments, the System should be observed to verify normal
operation in this setting.
The System communicates by low level RF energy. As with all RF
receivers, the potential for disturbance exists, even with equip-
ment that complies with FCC and CISPR emissions requirements.
The mylife OmniPod System communicates with the following
characteristics:
Frequency: 433 Mhz, FSK modulation, with an effective radi-
ated power of 13mW
The mylife OmniPod System greatly exceeds the immunity
requirements of the general standard for electromagnetic com-
patibility, IEC 60601-1-2.
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure
that is used in such an environment.
Emissions
Compliance according to
Electromagnetic environment
RF Emissions (CISPR 11)
Group 2
The Pod and the PDM emit low level electromagnetic energy (RF) in
order to communicate. Although unlikely, nearby electronic equip-
ment may be affected.
CISPR B
Emissions Classification
Class B
The System is suitable for use in all establishments including domestic
establishments.
Electromagnetic Immunity
The System is intended for use in the electromagnetic environment specified below. You should observe these requirements in the
use of the System.
Immunity against
IEC 60601-1-2
test level
Compliance level
(of this device)
Electromagnetic environment
ElectroStatic Discharge,
ESD (IEC 61000-4-2)
contact discharge:
± 6 kV
air discharge: ± 8 kV
± 8 kV
± 15 kV
If floors are covered with synthetic material, try to avoid electro-
static discharges.