Alerts and Alarms
129
10
PDM Error.
Memory corruption.
Press “
OK
” to reset PDM
and delete all user set-
tings. History records will
not be deleted.
Hazard
Alarm
PDM
Continuous tone.
This alarm occurs after the PDM is reset
and a memory corruption error occurs.
Press
OK
to acknowledge.
Change Pod.
Check blood glucose.
System Error.
Remove pod now.
For customer support,
refer to your User Guide.
Hazard
Alarm
Pod
or
PDM
Continuous tone.
This alarm occurs if an unexpected error
is detected in the Pod or PDM. Press
OK
to acknowledge.
Remove Pod.
Call Customer Care immediately.
Check blood glucose.
PDM Error.
Press
OK
to reset clock.
Hazard
Alarm
PDM
Continuous tone.
This alarm occurs after the PDM is reset,
indicating that the time and date must
be reset. Press
OK
to acknowledge.
Reset clock.
Check blood glucose.
Alarm
Type
Source
Beep Pattern
Action