Alerts and Alarms
125
10
If you ever need to return the PDM for repair or
replacement, contact your healthcare provider
for instructions on going back to treatment by
injections.
Use the form at the end of this
User Guide
to write
down all your settings from the Setup Wizard. If you
ever need to reset your PDM, having these settings
handy will make setup very easy.
Alarm
Type
Source
Beep Pattern
Action
End of insulin suspend.
Press
OK
to resume basal
rate delivery.
Advisory
Alarm
Pod
and
PDM
2 sets of beeps every minute for 3
minutes.
Repeats every 15 minutes until
insulin delivery is resumed.
Press
OK
to resume insulin delivery at
programmed
basal rate
.
Auto-off alert
Advisory
Alarm
Pod
and
PDM
2 sets of beeps every minute for 15
minutes.
Escalates to
Auto-off
hazard alarm
if not acknowledged within 15
minutes.
Press
OK
to acknowledge.
Insulin delivery will not resume
until you press
OK
. If you do not
press
OK
to resume insulin deli-
very, you could develop hyper-
glycemia (high blood glucose).
The Auto-off alert will escalate to
a hazard alarm if ignored and
will result in the deactivation of
your active Pod. Be sure to
respond to the alert when it
occurs.