Appendix
148
■
Accuracy Test Results
The following graphs are designed to show flow accuracy of the
Pod against given time periods.
Flow rate from startup
The following graph shows changes in the flow rate over the
stabilization period. The measurements were made using a Pod
with a basal rate of 0.05 U/h at high operating temperature.
Flow rate error
The following trumpet curve shows the accuracy of the delivery
rate in relation to the observation period. The measurements
were made using a Pod with a basal rate of 0.05 U/h at high oper-
ating temperature. The overall mean percentage flow error was
1.40%.
Summary of Contents for OmniPod my life
Page 1: ......
Page 62: ...Using the Pod 49 5 Figure 5 13 Adult Front Back Figure 5 14 Youth Front Back ...
Page 186: ......