Vector and Bar Graph Displays
Item
Specifications
Vector display
Displays the phase differences of the fundamental signals using vectors. This feature can only be
used on models with the harmonic measurement option.
Bar graph display
Displays a bar graph of the amplitudes of each harmonic. This feature can only be used on models
with the harmonic measurement option.
Trend Display
Displays a graph of measurement function numeric data trends.
The trends of up to eight functions can be displayed.
Data Storage
Item
Specifications
Binary save
Saves data directly to the medium* at the data update interval. The maximum file size is 1 GB.
ASCII save
Automatically or manually converts data saved in binary format to ASCII data (.csv extension) after
measurement finishes. The maximum file size is 4 GB.
Internal memory
Approximately 20 MB
Store interval
(Waveform display off)
100 ms to 99 hours 59 minutes 59 seconds
Internal memory maximum storage time estimates (With waveform display and the integration feature off)
Measurement
Channel
Number of Measured Items
(On each channel)
Store Interval
Max. Storage Time
Channel 1
3
100 ms
Approx. 40 hours
Channel 1
10
1 s
Approx. 120 hours
Channel 3
10
100 ms
Approx. 4 hours
Channel 3
20
1 s
Approx. 20 hours
The maximum storage time may be lower than the values given above depending on factors
such as user-defined function and integration settings.
The USB direct save interval varies depending on the number of saves and the specifications of
the memory. The maximum file size is 1 GB.
* The available storage media are USB storage and the internal RAM disk.
Saving and Loading of Data
Setup parameters, numeric data, waveform display data, and screen image data can be saved to a storage medium.
*
Saved setup parameters can be loaded from the storage medium.
*
* The available storage media are USB storage and the internal RAM disk.
Harmonic Measurement
Item
Specifications
What is measured
All installed elements.
Method
PLL synchronization
Frequency range
The range for the fundamental frequency of the PLL source is 10 Hz to 1.2 kHz.
PLL source
• Can be set to an input element voltage or current or to an external clock signal.
• Input level
50% or more of the rated measurement range when the crest factor is 3.
100% or more of the rated measurement range when the crest factor is 6.
• The frequency filter must be turned on when the fundamental frequency is less than or equal to
440 Hz.
FFT data length
1024
FFT processing word length 32 bits
Window function
Rectangular
Anti-aliasing filter
Set using the line filter. (Set to OFF or 5.5 kHz.)
14.6 Functions