1-7
IM 760201-01E
Component Names and Functions
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Other Keys
SETUP
INPUT INFO
DISPLAY
NUMERIC
WAVE
OTHERS
FORM
ITEM
CURSOR
INTEGRATOR
START/
STOP
RESET
HOLD
SINGLE
SHIFT
MISC
NULL
LOCAL
KEY LOCK
START/STOP Key
Use this key to start or end integration. You can view the integration condition in the
integration setup/condition display (see section 1.3).
SHIFT+START/STOP (RESET) Key Combination
Use this key combination to reset the integrated value.
HOLD (See section 4.12)
Press HOLD to stop data measurement and display operations and to hold the numeric
display (the HOLD key illuminates when the numeric display is held). During integration,
the numeric display is held but measurement is not stopped. Press HOLD again to allow
the numeric data display to be updated (the HOLD key will no longer be illuminated).
SHIFT+HOLD (SINGLE) Key Combination (See section 4.12)
Use this key combination while the display is held to measure the signal once at the set
data update rate, and then to re-hold the display.
LOCAL
Use this key to switch from remote mode (the REMOTE indicator will appear at the upper
right of the screen) to local mode (in which front panel key operations are valid). This key
is invalid when the WT500 is in local lockout mode.
SHIFT+LOCAL (KEY LOCK) Key Combination
Use this key combination to turn the key lock on or off. When the key lock is on, no keys
or key combinations other than SHIFT+LOCAL are valid, and LOCK appears in the upper
right of the screen.
MISC
Use this key to display a menu for viewing the system status, initializing setup
parameters, configuring or viewing the settings of remote control through communication
commands (using the USB, GP-IB, or Ethernet interface option), setting the date/time,
selecting the message and menu languages, selecting the crest factor, setting the USB
peripheral interface, setting the optional Ethernet interface, setting the self test, etc.
SHIFT+MISC (NULL) Key Combination
Use this key combination to activate the NULL feature, which removes the DC
component from the sampled data (the NULL indicator appears on the screen when the
NULL feature is activated). Press SHIFT+MISC (NULL) again to deactivate the NULL
feature (the NULL indicator will no longer appear).
SHIFT
When you press the SHIFT key, the key illuminates, and pressing a panel key produces
the effect indicated by the purple letters below it. Holding down SHIFT for 2 seconds or
more will lock the WT500 into the shifted state. Pressing SHIFT again will release the
shifted state (and the SHIFT key will no longer be illuminated).
1.2 Setup Menu Display and Operation Keys