![YOKOGAWA WT500 User Manual Download Page 319](http://html.mh-extra.com/html/yokogawa/wt500/wt500_user-manual_911418319.webp)
14-18
IM 760201-01E
14.14 General Specifications
Item
Specifications
Warm-up time
Approx. 30 minutes
Operating environment
Temperature: 5 to 40°C
Humidity: 20 to 80%RH
(No condensation)
Elevation
2000 m or less
Location
Indoors
Storage environment
Temperature: –25 to 60°C
Humidity: 20 to 80%RH
(No condensation)
Rated supply voltage
100 to 240 VAC
Permitted supply voltage range
90 to 264 VAC
Rated supply frequency
50/60 Hz
Permitted supply voltage frequency range 48 to 63 Hz
Maximum power consumption
250 VA
External dimensions
Approx. 213 mm × 177 mm × 409 mm (W × H × D)
(For details, see section 14.15.)
(Excluding the handle and other protruding parts.)
Weight
Approx. 7 kg (the weight of the main unit with three input elements and all options
installed)
Installation position
Horizontal (a stand can be used). Do not install vertically or stack.
Battery backup
Setup parameters and the internal time clock are backed up with a lithium battery.
Safety standard
1
Standards EN61010-1
The overvoltage category (Installation category) is CAT II.
2
The measurement category is CAT II.
3
The pollution degree is 2.
4
Emissions
1
Standards
EN61326-1 Class A
EN55011 Class A, Group 1
EN61000-3-2
EN61000-3-3
This is a class A instrument designed for an industrial
environment. Operation of this equipment in a residential
area can cause radio interference, in which case users will be
required to correct the interference.
C-Tick EN55011 Class A, Group 1
Cable conditions
• Ext Clk terminal
Use a BNC cable.
5
• GP-IB interface connector
Use a shielded GP-IB cable.
5
• RGB video signal output (VGA) connector
Use a shielded D-sub 15 pin VGA cable.
5
• USB port (PC)
Use a shielded USB cable.
5
• USB port (for peripheral devices)
Use a USB keyboard that has a shielded cable.
• Ethernet connector
Use a category 5 or better Ethernet cable (STP).
Immunity
1
Standards
EN61326-1 Table 2 (Industrial Environment)
Cable conditions
The same as the cable conditions listed above for emissions.
1 Applies to products with CE marks. For information concerning other products, contact your nearest YOKOGAWA dealer.
2 The overvoltage category is a value used to define the transient overvoltage condition and includes the rated impulse
withstand voltage. CAT II applies to electrical equipment that is powered through a fixed installation, such as a wall outlet wired
to a distribution board.
3 Measurement Category II (CAT II) applies to electrical equipment that is powered through a fixed installation, such as a wall
outlet wired to a distribution board, and to measurement performed on such wiring.
4 The pollution degree refers to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution degree 1 applies to closed atmospheres (with no, or only dry, non-conductive pollution). Pollution degree 2
applies to normal indoor atmospheres (with only non-conductive pollution).
5 Use cables that are 3 m or shorter.