![YOKOGAWA WT500 User Manual Download Page 87](http://html.mh-extra.com/html/yokogawa/wt500/wt500_user-manual_911418087.webp)
3-26
IM 760201-01E
3.13 Initializing the Settings
Procedure
Note
Only initialize the WT500 if you are sure that it is okay for all of the settings to return to their
initial values. You cannot undo an initialization. We recommend that you save the setup
parameters before initializing the WT500 (see section 10.3 for details).
1.
Press
MISC
to display the Misc Menu.
2.
Use the
cursor keys
to select Initialize Settings.
3.
Press
SET
. An alert message appears.
4.
Use the
cursor keys
to select OK.
5.
Press
SET
to execute initialization.
To cancel the initialization, select Cancel in step 3 and press SET.
Explanation
The values specified by using the panel keys can be reset to their factory default values.
This is useful when you wish to clear previous settings or start the settings from scratch.
For information about the initial settings, see appendix 2, “List of Initial Settings and
Numeric Data Display Order.”
Settings That Cannot Be Initialized
• Date and time settings
• Menu and message language settings
• The designation of the files display (Filter)
• File utility operation (Function)
• USB interface, GP-IB interface (optional), and Ethernet interface (optional) settings
Initializing at Power Up
To make the WT500 start up with the initial settings, turn on the power. Then, when the
message “Boot OK” appears, press and hold ESC until the NUMERIC key blinks. All of
the settings listed under “Settings That Cannot be Initialized,” except for the date and
time, are initialized. A message indicating that the settings have been initialized appears
on the screen.