6-17
IM 760201-01E
Harmonic Measurement (Optional)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Measuring with Cursors
4.
Press
SHIFT+FORM(CURSOR)
to display the Cursor menu.
• Selecting Whether or Not to Use Cursor Measurement
5.
Use the
cursor keys
to select Bar Cursor.
6.
Press
SET
to select ON or OFF. If you select ON, the results of cursor
measurement are displayed.
• Moving the Cursors
7.
Use the
cursor keys
to select Position C1 + or Position C2 x.
8.
Press
SET
to display the cursor position selection box for Position C1 + or
Position C2 x.
9.
Use the
cursor keys
to set the position in terms of the harmonic order.
• Select Position C1 + to move the position of .
• Select Position C2 x to move the position of cursor x.
10.
Press
SET
or
ESC
to close the cursor position selection box.
• Linking Cursors
11.
Use the
cursor keys
to select Linkage.
12.
Press
SET
. The linkage indicator lights and the cursor settings are linked.
When Linkage is enabled, the distance between and cursor x is maintained. When
you change the position of one cursor, the position of the other cursor also changes.
Linkage indicator
6.7 Displaying Bar Graphs and Making Cursor Measurements