6-1
IM 704610-01E
Storing and Recalling Setup Information
6
Chapter 6
Storing and Recalling Setup Information
6.1
Storing Setup Information
Procedure
MEDIA
FUNCTION
POLARITY
ARMING
INHIBIT
LOCAL
RECALL
ENTER
TRIG
IMPEDANCE EQUALIZER
SEC RANGE
KEY
LOCK
MEASURE
DISP OFF
JUDGE
GO
NO-GO
10%
20%
S
SEC RANGE
SCALE
L.MARK
R.MARK
GATE TIME
DELAY
BLOCK
REMOTE
STORE
UTILITY
JDG LEVEL
LEVEL
AUTO
COUPLE
EQ BOOST
MAN
CONV
50
Ω
1M
Ω
PW
E2T
D to C
BD x1
ns ms
s
%
V
%/
S
SHIFT
SHIFT
Display 1
Display 2
The phrase, “rotary knob & <>”
is used in the following
explanation to instruct the user
to enter numerical setting values
using the knob and arrow keys.
For details on this procedure,
see section 3.7.
Selecting the Memory Number
1.
Press SHIFT+RECALL (STORE). Store appears in Display 2, and the memory
number of the store destination appears in Display 1.
2.
Turn the rotary knob to select the memory number.
Executing the Store
3.
Press ENTER. The word donE is displayed for approximately one second, and
the setup information is stored. When the storing operation is finished, the
display once again shows measured results.
Explanation
Information That Is Stored
All settings other than communication related settings are stored to the internal memory.
Number of Sets That Can Be Stored
The number of sets that can be stored (memory numbers) is seven, from 0 to 6.
Note
If you initialize all setup information of the TA220 to factory default settings (see section 11.2),
the stored setup information is also initialized. The setup information of all seven sets is reset
to the factory default condition.