12-15
IM 704610-01E
Troubleshooting and
Maintenance
12
Pulse Width Jitter Measurement Test
Items Required
The following items are required.
Synthesized signal generator
• Frequency range: 1 MHz to 20 MHz
• Output level: 720 mVrms or more
• Recommended instrument: Synthesized signal generator 8648A (Agilent
Technologies)
The procedure for testing the pulse width jitter measurement using the
recommended instruments is described below.
Device Connections
• Ensure that all instruments are turned OFF before making any connections.
• Connect the output terminal (OUTPUT) of the synthesized signal generator to the
measurement input terminal (RF input) on the TA220.
Synthesized signal generator
(8648A)
TA220
RF input
OUTPUT
Instrument Settings
• TA220
• Media: BD
• Measurement function: PW
• Polarity:
,
• Pulse width lower limit: 0.00 ns
• Pulse width upper limit: 999.90 ns
• Input impedance: 50
Ω
• Input coupling: DC
• Equalizer: OFF
• Trigger mode: MAN
• Slice level: 0.000 V
• Gate Time: 1000 ms
• AGC: OFF
• 8648A
• Output level: 360 mV
rms
• Output frequency: 1.0 MHz, 4.0 MHz, 8.0 MHz, 16.5 MHz
12.6 Performance Test