10-29
IM 704610-01E
Communication Command
10
10.2 Commands
:SAMPle:INHibit:POLarity
Function
Sets the polarity of inhibit or queries the current
setting.
Syntax
:SAMPLe:INHibit:POLarity {POSitive|
NEGative}
:SAMPle:INHibit:POLarity?
Example
:SAMPLE:INHIBIT:POLARITY OFF
:SAMPLE:INHIBIT:POLARITY?
-> :SAMPLE:INHIBIT:POLARITY OFF
Description Cannot be set or queried when
SAMPle:INHibit:STATe (inhibit) is OFF.
:SAMPle:INHibit:STATe
Function
Turns ON/OFF inhibit or queries the current
setting.
Syntax
:SAMPLe:INHibit:STATe {<Boolean>}
:SAMPle:INHibit:STATe?
Example
:SAMPLE:INHIBIT:STATE OFF
:SAMPLE:INHIBIT:STATE?
-> :SAMPLE:INHIBIT:STATE OFF
10.2.13 SSTart Group
:SSTart
:SSTart
Function
Executes single measurement.
Syntax
:SSTart
Example
:SSTART
Explanation Cannot be executed when
INPut:HISPeed:STATe (D-to-C high speed
calculation) is ON.
10.2.14 STARt Group
:STARt
10.2.15 STATus Group
This group consists of statistical computation related commands. See section 10.3 for more information on the status
report.
:STATus
:
CONDition
?
EESE
<Space>
<Register>
?
EESR
?
ERRor
?
FILTer
<x>
<Space>
RISE
FALL
BOTH
NEVer
?
QMESsage
<Space>
OFF
ON
<NRf>
?
?
;
:STARt
Function
Starts measurement.
Syntax
:STARt
Example
:START