4-13
IM 704610-01E
Measurement Condition Settings
4
Inhibit Signal
Input the inhibit signal to the connector labeled INHIBIT IN on the rear panel.
Item
Description
Connector type
BNC
Input Impedance
10 k
Ω
(typical value*)
Input coupling
DC
Input level
TTL Level
Allowable input voltage range
–5 to 10 V (DC+ACpeak)
Minimum input pulse width
30 ns
*
The typical value is a representative or standard value. It is not strictly guaranteed.
(Inhibit signal
input terminal)
Inhibit input circuit
IN
INHIBIT
(TTL)
+5 V
10 k
Ω
74LS14 or
equivalent
470
Ω
CAUTION
Do not apply a voltage that exceeds the allowable input voltage range to the
inhibit signal input terminal. This may cause damage to the TA220.