13-3
IM 704610-01E
13
Specifications
13.4 Display
Item
Specifications
Meter (needle)
Item
Jitter ratio, jitter
Jitter ratio display scale
Select 10% or 20%.
Jitter display time range
Select 0.5 ns, 1.0 ns, 5.0 ns, 10 ns, 50 ns, 0.1
µ
s, 0.5
µ
s, 1.0
µ
s, or 5.0
µ
s.
Display accuracy:
±
1.5% of FS
Display 1 (7-segment red LED)
Item
Jitter ratio, jitter, average value, T (measuring range of pulse width, or clock period), no. of
samples, boost amount
†
, delay time
†
, voltage level
Jitter ratio display range:
0 to 20%,
Display resolution: 0.01%
Jitter display range:
0 to 99.999 ns, Display resolution: 0.001 ns
Average display range:
0 to 999.99 ns, Display resolution: 0.01 ns
Voltage level display range: 0 to 9.999 V,
Display resolution: 0.001 V
Unit: ns, ms, s, %, V
Display 2 (dot matrix green LED)
Item
•
Setup menu (setting parameters)
•
Meaning of numbers displayed in Display 1
•
Time range types when selecting time range on the meter
•
Boost amount or delay time (when these are displayed, Display 1 shows the jitter ratio)
Display ON/OFF
The numerical value/alphabetical character display and units indicators on displays 1 and 2 can
be turned ON or OFF simultaneously.
† When pressing the MEASURE key to switch the numerical value display, the numerical value is displayed in Display 2.