12-22
IM 704610-01E
12.8 Replacing the Power Fuse
WARNING
• To prevent electric shock or fires, use a fuse of the specified rating in terms of
current, voltage, and type.
• Make sure to turn OFF the instrument and unplug the power cord before
replacing the fuse.
• Do not short the fuse holder.
• A fuse is placed on only one end of the power supply line.
Fuse Rating
The power supply fuse used by this instrument has the following specifications.
Maximum rated voltage
250 V
Maximum rated current
5 A
Type
Time lag
Specifications
UL/VDE
Part No.
A1114EF
Replacement Method
Replace the fuse according to the following steps.
1.
Turn OFF the power switch.
2.
Unplug the power cord from the power connector.
3.
Insert the tip of a flathead screwdriver into the indentation on the fuse holder
located near the power connector. Turn the driver in the direction of the arrow,
and remove the fuse holder.
4.
Remove the blown fuse from the tip of the fuse holder.
5.
Install a new fuse into the fuse holder, then reinsert the fuse holder into its
original location.
Fuse holder
Spare fuse
Fuse