12-20
IM 704610-01E
Test Procedure
• Test the TA220/FG300 after a thirty-minute warm-up.
• Input the offset modulated FG signal to the TA220, and confirm that the standard
deviation
σ
(jitter) of the pulse width jitter measurement is within the allowable range.
The average value is the reference value.
Test procedure
1.
Set the modulation frequency of the FG300 according to the values in the table
below.
2.
Measure the pulse width of the modulated signal on the TA220, and confirm that
the standard deviation
σ
(jitter) is within the allowable range. The average value
is the reference value.
Test Results
FG modulation frequency
Standard deviation
σ
(jitter)
Average
Measured
Judgment
Measured
Reference
value
criteria
value
value
1.0 Hz
1.0 ns or less
62.5
±
1 ns
1.0 kHz
1.0 ns or less
62.5
±
1 ns
5.0 kHz
20 ns or less
62.5
±
1 ns
10.0 kHz
60 ns or less
62.5
±
1 ns
12.6 Performance Test