App-8
IM 767361-01E
OM
Requests output of the number of data items it
is possible to store.
Syntax
OM <terminator>
Response Example
FREE2000 <terminator>
ORD
Requests output of the recall data.
Syntax
ORDm, n <terminator>
“m” represents the number of the first
memory location to be recalled.
m=1 to 9999
“n” represents the number of the last
memory location to be recalled.
n=1 to 9999(n≥m)
Description
• Cannot execute unless in recall mode.
An error 15 occurs.
• 100 data can be output at one time.
If a memory data higher than 100 is
specified, only the first 100 data will be
output.
• If the end memory number is smaller
than the start memory number, error 12
occurs.
• Refer to page App-15, “Output Format
for Recall Data.”
OS
Requests output of the setting information.
Syntax
OS<terminator>
Description
Refer to page App-15, “Output Format for
Panel Setting Information.”
PD/PD?
Sets the number of display digits for the
measured pressure value and the value to be
displayed in %, and queries the current setting.
Syntax
PDm<terminator>
“m” represents the number of display
digits.
0: Measured pressure value in 5-1/2
digits, % value in 4-1/2 digits
1: Measured pressure value in 4-1/2
digits, % value in 3-1/2 digits
Query
PD?<terminator>
Response Example
PD0
Description
In the case of 700 kPa models
(767355/767365/767373/767385), “m”
represents the number of display digits
as follows:
0: Measured pressure value in 4-1/2
digits, % value in 4-1/2 digits
1: Measured pressure value in 3-1/2
digits, % value in 3-1/2 digits
PU/PU?
Sets the units of pressure and queries the
current setting.
Syntax
PUm<terminator>
The value “m” is the unit of pressure.
1: inH
2
0
(available with the -U3 model)
2: inHg
(available with the -U3 model)
3: psi
(available with the -U3 model)
4: kPa
5: kgf/cm
2
(available with the -U2 and
-U3 models)
6: mmH
2
0 (available with the -U2 and
-U3 models)
7: mmHg (available with the -U2 and
-U3 models)
Query
PU?<terminator>
Response Example
PU4
Appendix