Overview of the RS-232C Interface
59
IM 767361-01E
Overview of the RS-232 Interface
RS-232 Communications Settings
Use the settings below to setup communications so that you can enter settings
on the instrument using the PC, or output setting information or values to the
computer.
• Selecting the Handshaking Method
Choose a data sending/receiving control from the following:
Settings
Handshaking Method
0 OFF-OFF
1 XON-XON
2 XON-RTS
3 CTS-RTS
• Setting the Data Format
Choose a data length–parity–stop-bit combination from the following:
Settings
Data Length
Parity bit
Stop bit
0 8 none
1
1 7 ODD
1
2 7 EVEN
1
3 7 none
2
• Setting the Baud Rate
Choose from the options below:
1200
2400
4800
9600
Selecting Normal or Talk-Only Mode
The RS-232C interface for this instrument has two control modes: normal mode
and talk-only mode. Normal mode is equivalent to the addressable mode for the
GP-IB interface, and enables reception of commands and transmission of data.
Measured or computed data is output on receipt of the “OD” command.
Talk-only mode is equivalent to the talk-only mode for the GP-IB interface. Only
transmission of measured or computed data is allowed. No commands can be
received.
1.
Press
MENU
.
Keep pressing or until
is displayed in the screen, then press
ENTER/LOCAL
.
2.
Press or to select
(normal mode) or
(talk-only mode),
then press
ENTER/LOCAL
.
3.
Press
MENU
to return to the normal measurement screen. However, if you want
to set the handshake mode, data format, or baud rate settings, proceed to “Setting
the Handshaking Mode, Data Format, and Baud Rate” on this page.