Before Starting Measurement
23
IM 767361-01E
Before Starting Measurement
Power Rating
Rated supply voltage range
: 100 to 120 VAC/220 to 240 VAC
Permitted supply voltage range
: 90 to 132 VAC/180 to 264 VAC
Rated supply voltage frequency
: 50/60 Hz
Permitted supply voltage frequency range : 47 to 63 Hz
Power consumption (pressure measurement only)
: 25 VA MAX (100 VAC)
: 40 VA MAX (200 VAC)
Power consumption (pressure measu DMM + 24V charging
time)
: 45 VA MAX (100 VAC)
: 65 VA MAX (200 VAC)
Supplying DC Power
Connecting the Power Cord
The pin assignment of the external DC power connector is shown below. Use a
cord that matches the DC power to be used.
(
-
)
(
+
)
1
3
2
(
-
)
Note
Pins 1 and 2 are short-circuited internally.
Power Rating
Rated supply voltage range
: 10 to 15 VDC
Permitted supply voltage range : 9 to 16.5 VDC
Maximum power consumption : 10 VA MAX
WARNING
• To prevent electric shock, use a power supply not exceeding 60 VDC.
CAUTION
• Never short-circuit the terminals.
• If a DC voltage exceeding the permitted supply voltage range is input to
the instrument, damage may result.
• The number 1 and 2 connector pins of the external DC power supply
have the same electric potential as the meter case. Do not connect a
floating power supply.
• Do not supply AC power when a DC power supply is used. It may
damage the instrument.