Appendix
App-9
IM 767361-01E
Description
• The available settings vary depending
on the pressure units specification code
you selected at the time of purchase.
• The units cannot be changed when in
% or %ERROR display mode. An error
15 occurs.
PZ
Performs zero calibration for pressure
measurement.
Syntax
PZ<terminator>
Description
• As shown below, if the reference
pressure value is wrong, zero
calibration cannot be performed. An
error 17 occurs.
• An overrange occurred in the
measured pressure value (and “--OL-
-” was displayed).
• When using an absolute pressure
model (767357/767367/767387), the
value differs from the factory default
by more than 1 kPa.
• Cannot execute in recall mode. An
error 14 occurs.
RC
Initializes the setup information.
Syntax
RC<terminator>
REL/REL?
Sets the relative display and queries the current
setting.
Syntax
RELm<terminator>
“m” represents the display setting.
0: OFF
1: ON
Query
REL?<terminator>
Response Example
REL0
Description
• When an overrange occurred in the
measured pressure value (and “--OL-
-” was displayed), the relative display
could not be turned ON. An error 15
occurs.
• The relative value cannot be displayed
when in % or %ERROR display mode.
An error 15 occurs.
• Cannot change in recall mode. An
error 14 occurs.
RNO/RNO?
Sets the memory Number to be recalled/queries
the current setting.
Syntax
RNO m <terminator>
“m” represents the memory Number to be
recalled.
m=1 to 9999
Description
Cannot set unless in recall mode. An
error 15 occurs.
RO/RO?
Turns recall mode ON and OFF/queries the
current setting.
Syntax
RO m <terminator>
“m” represents ON/OFF for recall mode.
0: Recall mode is OFF
1: Recall mode is ON
Query
RO? <terminator>
Response Example
RO0
SI/SI?
Sets the sampling rate and queries the current
setting.
Syntax
SIm<terminator>
“m” represents the sample rate.
0: 250 ms
1: 4 s
Query
SI?<terminator>
Appendix