Description
• Only valid with the D/A output option (/
DA)
• The parameters are set in the order
lower limit, then upper limit. The
setting, Lower limit ≥ upper limit, is not
possible. An error 12 occurs.
CMP/CMP?
Turns comparator output ON/OFF, and queries
the current setting.
Syntax
CMPm<terminator>
“m” represents the comparator output
0: OFF
1: ON
Query
CMP?<terminator>
Response Example
CMP1
Description
Only valid with the D/A output option (/
DA)
CVD
Sets the value of the DC voltage (5-V
range) in the calibration point and queries
the measurement value of the input before
calibration.
Syntax
CVDm<terminator>
“m” represents the calibration value.
0 (CVP1):
0.0000 <fixed>
+FS(CVP2): 0.00000 to 6.00000
-FS(CVP3): -6.00000 to 0.00000
Response Example
V_N_5.0000
Description
• Error 11 occurs if the mode is not DMM
calibration mode.
• If the item to be calibrated is not set to
DC voltage (CAL1), an error 21 occurs.
• For the calibration value for calibration
point 0 (CVP1), set m to 0.
• For details regarding the calibration
procedure, refer to pages 57 and App-
18.
CVP/CVP?
Sets the calibration points for the DC voltage (5
V range), and queries the current setting.
Syntax
CVPm<terminator>
“m” represents the calibration point.
0: After calibration, the correction
coefficient is calculated, then the result
is written to the EEPROM.
1: 0(0V)
2: +FS (5 V)
3: -FS (-5 V)
Query
CVP?<terminator>
Response Example
CVP1
Description
• Error 11 occurs if the mode is not DMM
calibration mode.
• If the item to be calibrated is not set to
DC voltage (CAL1), an error 21 occurs.
• If the three calibration points have not
been set when end (end of calibration)
is selected, calibration error 21 occurs.
Also, if the calculated correction
coefficient is not correct, calibration
error 22 occurs.
• For details regarding the calibration
procedure, refer to pages 57 and App-
18.
DA
Deletes all the stored data.
Syntax
DA <terminator>
Description
Cannot be executed during auto store.
An error 14 occurs.
DB/DB?
Deletes the specified stored data.
Appendix