22
IM 767361-01E
Before Starting Measurement
Supplying AC Power
Connecting the Power Cord
Make sure that you perform the following steps before connecting the
power. Failure to do so may cause an electric shock or cause damage to the
instrument.
WARNING
• Before connecting the power cord, ensure that the source voltage
matches the rated supply voltage of the instrument and that it is within
the maximum rated voltage of the provided power cord.
• Connect the power cord after confirming that both the MAIN POWER
switch (rear panel) and the POWER switch (front panel) are OFF.
• To prevent electric shock or fire, use only the power cord supplied with
this instrument.
• Make sure to connect protective earth grounding to prevent electric
shock. Connect the power cord to a three-prong power outlet with a
protective earth terminal.
• Do not use an extension cord without protective earth ground. The
protective features of the instrument will be rendered ineffective.
French
AVERTISSEMENT
• Avant de brancher le cordon d’alimentation, vérifier que la tension
source correspond à la tension d’alimentation nominale du DLM4000
et qu’elle est compatible avec la tension nominale maximale du cordon
d’alimentation.
• Brancher le cordon d’alimentation après avoir vérifié que l’interrupteur
d’alimentation principal de l’instrument est sur OFF.
• Pour éviter tout risque de choc électrique ou d’incendie, utiliser
exclusivement le cordon d’alimentation fourni par YOKOGAWA et prévu
pour l’instrument.
• Relier l’instrument à la terre pour éviter tout risque de choc électrique.
Brancher le cordon d’alimentation sur une prise de courant à trois plots
reliée à la terre.
• Toujours utiliser une rallonge avec broche de mise à la terre, à défaut de
quoi l’instrument ne serait pas relié à la terre.