Appendix
App-15
IM 767361-01E
Output Format for Recalled Data (Response to the “ORD” Command, MT220
Only)
Data Number
h1 h2 h3 h4 d1 d2 d3 d4
h1-h3
NO.: Recall data number
h4
: Recall data is present.
E: No recall data is present.
d1-d4
4 digit number (recall data number)
Output Format
Pressure
data
Pressure
header
,
Data
number
,
DMM data
DMM header
Terminator
Output Examples
• DMM function OFF, normal display of recall data
NO. 0001, KPAN 130.000, V N 05.0000
• No recall data
NO.E0002, KPAE 888888., V E 888888.
Output Format of Setup Information (Response to the “OS” Command)
For the MT210/MT210F
1st line: Model, software version number.
2nd line: Pressure setup information
3rd line: Measurement condition
4th line: Backlight and beep
5th line
*1
: D/A output and comparator settings
6th line: Output end
*1: Line 5 is output only when the D/A output option (/DA) is installed.
Example of response
MDL767313;REV1.01
REL0;PU4;PD0
HD0;AG1
BL0;BP1
DR0;CMP0;CMD0.000,130.000
END
For the MT220
1st line: Model, software version number
2nd line: Pressure set-up information
3rd line: DMM set-up information
4th line: Measurement conditions
5th line: Power output, back light, beep
6th line: Data memory function
7th line
*2
: D/A output and comparator settings
8th line: Output end
*2: 7th line output only with the D/A output option (/DA) installed
Appendix