34
IM 767361-01E
• Other than 767360
Mesurement Mode
Normal
Medium-speed
High-speed
Pressure Averaging
ON
OFF
Minumum Trigger Input Interval
MT210/MT210F MT220
1.6 s
260 ms
110 ms
80 ms
DMM ON
DMM Averaging
ON
2.5 s
2.5 s
OFF
1.6 s
500 ms
DMM OFF
1.6 s
500 ms
• 767370
Measurement Mode
Sample Rate
Averaging
Minimum Trigger Input Interval
Normal
4000 ms
–
4.1 s
250 ms
ON
2.6 s
OFF
260 ms
D/A Output (/DA Option)
Converts pressure measurement results from digital to analog and outputs
them. When starting the instrument using DC power or the battery pack, the D/A
output function is turned OFF. Turn it ON before using the instrument.
• Output range
: 2 V or 5 V
• D/A conversion time : 2 ms
• The relationship between pressure measurement values and the output
voltage (for the 130-kPa range)
Output Voltage
Measured Pressure (kPa)
130.000
5-V range
2-V range
–1.231
–3.077
5.000
2.000
–80.000
0
Pressure Measurement