Troubleshooting, Maintenance, and
Inspection
63
IM 767361-01E
Troubleshooting, Maintenance, and Inspection
No. Type of Error
Cause
Corrective Action
Page
33
Store error
Internal memory full
Delete unneeded measurement data.
36 to 38
38
Store error
Attempted to store measurement data
zunder an already existing data number.
Delete the existing data, or use an
unused data number.
50
Error in measured
pressure value
Measured pressure value update error.
“---------” appears in the upper part of the
screen.
Service required if this continues to be
displayed.
–
51
Error in measured
pressure value
Overrange occurred in the measured
pressure value. “--oL--” is displayed in the
upper part of the screen.
Input a pressure within 120% of the
measurement range.
24
52
Error in measured
pressure value
Overflow occurred in the measured
pressure value.
“--oF--” is displayed in the upper part of
the screen (or [SPn.Err] when in % display
mode).
If [SPn.Err] is displayed, reenter the
correct pressure value for 0%, 100%.
34, 35
53
DMM measurement error DMM A/D conversion timeout.
Service required.
–
54
DMM measurement error DMM measurement update error.
“---------” appears in the lower part of the
screen,
Service required if this continues to be
displayed. (Service is required if, when
using the OE communications command
to readout the error message, that
message to be displayed.)
–
55
DMM measurement error DMM measurement overrange.
“--oL--” appears in the lower part of the
screen.
Input a pressure value within the
displayed range.
34, 35
56
DMM measurement error
%ERROR calculation overflow.
“--oF--” is displayed in the lower part of the
screen.
Check whether the pressure percentage,
voltage, and current values are all
correctly displayed.
34, 35
Self-Test Error When the Power is Turned ON.
No. Type of Error
Cause
Corrective Action
Page
60 Setting information
backup error (excluding
communications settings)
Panel setting information other than
communications settings were corrupted.
Initialize the settings.
22
61 Communications settings
information backup error
File is damaged.
Initialize the settings.
22
62 Measurement data backup
error
Measurement data saved to the internal Initialize
the data.
memory was corrupted.
–
63 Zero calibration backup value
failure (absolute pressure
models only)
Values for Zero CAL were corrupted.
(767357/767367/767387)
Initialize the Zero CAL values. 23
83 EEPROM error (DMM
calibration)
EEPROM values were corrupted.
DMM calibration is required.
56 to 58
84 EEPROM error (D/A output
calibration value)
EEPROM values were corrupted.
Service is required.
–
90 Hardware error
Hardware failure in the pressure measurement
section.
Service required.
–
91 Pressure measurement
section calibration data
failure (sensor section)
Calibration data from the pressure sensor were
corrupted.
Service required.
–
92 Calibration error in the
pressure measurement
(calculation) section
Calibration data from the pressure calculation
section were corrupted.
Service required.
–
93 Setting value error in the
pressure measurement
section (sensor section)
Setting data from the pressure sensor were
corrupted.
Service required.
–
94 Setting value error in the
pressure measurement
(calculation) section
Failure in the setting values of the pressure
calculation section.
Service required.
–
• If an error 60–63, 83, or 84 occur when turning ON the power, the error will be displayed until a certain key is pressed.