Troubleshooting, Maintenance, and
Inspection
67
IM 767361-01E
Troubleshooting, Maintenance, and Inspection
Fuse Ratings
When you order a spare fuse, specify the following part number.
Rating
Part Number
Location for Use
Replacement Procedure
250 V, 100 mA, time lag
A1341EF
Current input terminal See next page
250 V, 3.15 A, time lag
A1113EF
Rear panel
See next page
250 V, 2 A, time lag
A1429EF
Power supply board
Service required*
*: Professional service is required. Contact your nearest YOKOGAWA representative as listed on
the back cover of this manual.
Fuse Position and Replacement Method
Turn the fuse holder to the left and pull out the fuse holder. Then, replace with a
new fuse.
Turn the fuse holder to the left and pull out the
fuse holder. Then, replace with a new fuse.
Each current input terminal accommodates a built-in fuse. A fuse is contained
in the fuse holder as shown below. To replace the fuse, turn the fuse holder
counter-clockwise with your fingers to remove it, then replace the fuse with a
new one.