Specifications
73
IM 767361-01E
Specifications
Common Specifications
Display
LCD (with back lighting), pressure measurement section: 5-1/2 digits/4-1/2 digits
*1
DC V,DC A
measurement section: 4-1/2 digits
Warmup Time
Approx. 5 minutes
Operating Temperature/
5 to 40‚°C
*2
, 20 to 80% RH (no condensation allowed)
Humidity Range
Operating altitude
2000 m or less
Storage Temperature
–20 to 60°C
Range
Over Voltage Category
CAT II (EN61010-1)
Power Source
AC, DC, or Ni-Cd batteries (sold separately)
Rated AC Voltage
100 to 120 VAC/200 to 240 VAC
Rated AC Frequency
50/60 Hz
Permitted AC Voltage Range 90 to 132 VAC/180 to 264 VAC
Permitted AC Frequency
47 to 63 Hz
Range
Rated DC Voltage
10 to 15 VDC
Permitted DC Voltage Range 9 to 16.5 VDC
Battery Pack
External Ni-Cd batteries: For the MT210/MT210F
(Optional Accessory)
The instrument can be operated continuously for approximately 10
hours (with back light on) if the batteries have been charged fully.
For the MT220
The instrument can be operated continuously for approximately 6
hours (with back light, DMM, and 24 VDC output ON) if the batteries
have been charged fully.
However, this varies according to the operating conditions.
Charger: Built in to the MT120/MT210F/MT220.
Charging time: Approximately 12 hours
Maximum power consumption Pressure measurement: 25 VA max. (100 VAC), 40 VA max. (200 VAC)
Charging: 45 VA max. (100 VAC), 65 VA max. (200 VAC)
When operated on DC power: 10 VA max.
Insulation Resistance
More than 20 MΩ at 500 VDC (between AC power line and case)
Withstanding Voltage
1500 VAC for one minute at 50/60 Hz (between AC power line and case)
350 VAC for one minute at 50/60 Hz (between each DMM input terminal and 24 VDC output
terminal and case)
External Dimensions
Main unit approximately 132 × 213 × 350 mm (no projecting parts included)
Battery pack (optional) approximately 33 × 182 × 260 mm (no projecting parts included)
Weight
Refer to “Weight (Main Unit)” in the Pressure Measurement Section.
Battery pack (optional) approximately 2.7 kg
Accessories
Connector for external DC power (1), rubber feet (2), seal for measuring object, power cord (1),
and one user’s manual (this manual), Remote connector for the output terminal (when the /DA
option is installed) (1), Fuse (2) (1 is attached to the manometer), VCO body (1) (for -P3) or (2)
(for the 76737 -P3), fuse for current input terminal (2, one being pre-installed, MT220 only),
and one measurement lead (MT220 only).
*1 Values are displayed with 4-1/2 and 3-1/2 digits with the models 767355, 767365, 767373, and 767385.
*2 Operating temperature is 10 to 35°C (model 767370 only)