App-2
IM 767361-01E
Description
• The 100% value cannot be
automatically set when in % or
%ERROR display mode. An error 15
occurs.
• Cannot be executed while displaying
the relative value. An error 15 occurs.
• Cannot execute in recall mode. An
error 14 occurs.
AL
Sets the current measured pressure value as
the 0% value.
Syntax
AL <terminator>
Description
• The 0% value cannot be automatically
set when in % or %ERROR display
mode. An error 15 occurs.
• Cannot be executed while displaying
the relative value. An error 15 occurs.
• Cannot execute in recall mode. An
error 14 occurs.
BL/BL?
Turns the back lighting function ON and OFF,
and queries the current setting.
Syntax
BLm<terminator>
“m” represents ON/OFF for the back
lighting.
0: OFF
1: ON
Query
BL?<terminator>
Response Example
BL0
BP/BP?
Queries the setting for the beep sound (the
sound when keys are pressed).
Syntax
BPm<terminator>
“m” represents the ON/OFF for the beep
sound setting.
0: OFF
1: ON
Communications Command List
AG/AG?
Turns the averaging function ON and OFF, and
queries the current setting.
Syntax
AGm <terminator>
“m” represents ON/OFF for the averaging
function.
For the MT210/MT210F
0: Averaging is OFF for pressure
measurement.
1: Averaging is ON for pressure
measurement.
For the MT220
0: Pressure averaging OFF
DMM averaging OFF
1: Pressure averaging ON
DMM averaging OFF
2: Pressure averaging OFF
DMM averaging ON
3: Pressure averaging ON
DMM averaging ON
Query
AG? <terminator>
Response Example
AG1
Description
• For the MT210F, averaging settings
are only valid for normal measurement
mode. The averaging function does not
work when in medium and high-speed
measurement modes so it cannot be
selected. An error 15 occurs.
• Pressure averaging ON(AG1) cannot
be selected when the pressure
sampling rate is 4 s (SI1) on the
767370 (1 kPa differential model). An
error 12 occurs.
• Cannot change in recall mode. An
error 14 occurs.
AH
Sets the current measured pressure value as
the 100% value.
Syntax
AH <terminator>
Appendix