< PART – B Software >
B-87
IM 39J06B40-01EN
14th Edition: May 30, 2022
- For double-ended measurement, the Loss setting need not be specified because the measurement
results of the forward and reverse directions are combined. Even if specified, it will not affect the
temperature calculation result. However, the Loss setting is required when a fiber failure is
detected. If fiber failure happens, the measurement results up to the fiber failure position of the
forward and reverse directions are combined.
- Winding Coeff
Specify the winding coefficient of a fiber section as a value from 0.001 to 99.999 in
units of 0.001.
The winding coefficient is used when distance values are converted to depth values
for an axis using the following formula:
Depth = (Distance
-
Well Inlet Position)/ Winding Coeff.
SEE ALSO
For information on the setting parameters for calibration, correction and noise filter, see Chapter
B7.3.3, “Calibration”.
Summary of Contents for DTSXL
Page 2: ...Blank Page...
Page 15: ...Copyright and Trademark Notices xiii IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...
Page 23: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...
Page 83: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...
Page 215: ...PART B Software B 132 IM 39J06B40 01EN 14th Edition May 30 2022...
Page 222: ...PART B Software B 139 IM 39J06B40 01EN 14th Edition May 30 2022...
Page 233: ...PART B Software B 150 IM 39J06B40 01EN 14th Edition May 30 2022...
Page 239: ...PART B Software B 156 IM 39J06B40 01EN 14th Edition May 30 2022...
Page 247: ...PART B Software B 164 IM 39J06B40 01EN 14th Edition May 30 2022...
Page 253: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...
Page 289: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...