YOKOGAWA DTSXL User Manual Download Page 11

 

< Safety Precautions > 

ix

 

 

 

IM 39J06B40-01EN

 

14th Edition: May 30, 2022

 

 

 Laser Beam 

Safety Precautions for Laser Products 
This instrument uses a laser light source. This instrument is a Class 1M laser product as 

defined by IEC 60825-1:2007 Safety of Laser Products—Part1: Equipment classification 

and requirements, and a Class1 laser product as defined by IEC 60825-1:2014 Safety of 

Laser Products—Part1: Equipment classification and requirements. In addition, this 

instrument complies with 21 CFR Part 1040.10 except for deviations pursuant to Laser 

Notice No. 50, dated June 24, 2007. 
The table below shows the parameters of the laser light used in this instrument.  

Table   Parameters of Laser Light 

Laser Class 

Class 1M(2007)/ Class1 1(2014) 

Structure 

DFB-LD and EDFA 

Wavelength 

1552 nm 

Output power (min. and max.)  

3.8 W 

Pulse train 

Periodic pulse train 

Pulse time  

Max 10 ns 

Pulse repetition cycle 

Min 80 µs 

 
The Laser Class label is printed on the front panel of the instrument. 

 

 

Summary of Contents for DTSXL

Page 1: ...User s Manual IM 39J06B40 01EN DTSXL Distributed Temperature Sensor Long Range System Guide IM 39J06B40 01EN 14th Edition...

Page 2: ...Blank Page...

Page 3: ...e DTSXL System Before using the DTSXL System read this document and Read Me First IM 39J06A40 01EN Organization of this Manual This manual consists of the following chapters PART A Hardware This chapt...

Page 4: ...f prepare it separately Be sure to use the spare parts approved by Yokogawa Electric Corporation hereafter simply referred to as YOKOGAWA when replacing parts or consumables Modification of the produc...

Page 5: ...uals please contact our sales representative or your local distributor Warning and Disclaimer The product is provided on an as is basis YOKOGAWA shall have neither liability nor responsibility to any...

Page 6: ...tifies instructions that must be observed in order to avoid physical injury and electric shock or death of the operator CAUTION Identifies instructions that must be observed in order to prevent the so...

Page 7: ...EA European Economic Area are indicated below Manufacturer YOKOGAWA Electric Corporation 2 9 32 Nakacho Musashino shi Tokyo 180 8750 Japan Authorised representative in EEA Yokogawa Europe B V Euroweg...

Page 8: ...idity Operation 20 to 80 RH no condensation Installation height Altitude up to 3000 m 2 Installation category based on IEC 61010 3 Pollution degree based on IEC 61010 4 2 Installation Indoor use only...

Page 9: ...according to the procedure given in this document The power cables must conform to the safety standards of the country where the device is installed SEE ALSO For more information about power cables r...

Page 10: ...or DTSX3000 I O Devices WARNING To ensure this system compliance with the CSA safety standards all devices connected to this system shall be CSA certified devices Maintenance WARNING When the device b...

Page 11: ...r Products Part1 Equipment classification and requirements In addition this instrument complies with 21 CFR Part 1040 10 except for deviations pursuant to Laser Notice No 50 dated June 24 2007 The tab...

Page 12: ...To prevent accidents always observe the following precautions Do not turn on laser output when no optical fiber is connected to the optical connector Always stop laser output before disconnecting the...

Page 13: ...emperature installation environments Measurement Before performing distributed temperature measurement Ensure that the environment complies with the DTSXL System specifications and Perform temperature...

Page 14: ...r the like or registered or recorded on video tapes Trademark and License Software Acknowledgments DTSX and FAST TOOLS are registered trademarks of Yokogawa Electric Corporation STARDOM and CENTUM are...

Page 15: ...Copyright and Trademark Notices xiii IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...

Page 16: ...Supply Equipment A 16 A3 1 Power Supply System A 16 A3 2 Supplied Power Source A 16 A3 3 Grounding for DTSXL System A 18 A4 Installation and Attachment A 19 A4 1 Installing the Base Module A 19 A4 2 I...

Page 17: ...1 Functional Overview of the DTSX3000 B 1 B1 2 Functions B 2 B1 3 Connection B 6 B1 3 1 Connecting to Control Devices B 6 B1 3 2 Connecting to Various Software for DTSX3000 B 7 B1 4 IP Address Config...

Page 18: ...t User B 43 B6 4 4 Network Configuration Network B 46 B6 4 5 Date and Time Configuration Date B 48 B6 4 6 Web Server Configuration Web B 55 B6 4 7 SSH SFTP Configuration SSH SFTP B 57 B6 4 8 Serial Po...

Page 19: ...1 Specifications of Memory Area B 99 B8 3 2 File Name Structure B 100 B8 3 3 Archive Directory B 102 B8 4 Retrieving Measurement Data B 103 B8 4 1 Retrieving Files Using SFTP Client B 103 B8 4 2 HTTP...

Page 20: ...f Procedures and Operation C 1 C1 1 1 Power on Procedure C 1 C1 1 2 Power off Procedure C 1 C1 2 Reboot C 3 C1 3 System Operation upon Power Failure C 3 C2 DTS specific Operation and Precautions C 4 C...

Page 21: ...30 2022 C4 3 1 Batteries for DTSX3000 C 19 C4 3 2 Power Supply Module C 26 C4 3 3 Replacing Other Parts C 26 C4 4 Sensor Optical Fiber C 29 C4 5 Fault Isolation and Self Diagnosis C 30 C4 5 1 Self dia...

Page 22: ...May 30 2022 PART A Hardware This chapter describes the hardware overview of theDTSXL Distributed Temperature Sensor Long Range System DTSXL System DTSXL Distributed Temperature Sensor Long Range Syste...

Page 23: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...

Page 24: ...ch module module with built in optical switch required for channel switching in multi channel measurement Rack Mount kit required for reducing the load on the optical fiber connector Optical Fiber for...

Page 25: ...d on 19 inch Rack Code Module DTS DTSX3000 Distributed Temperature Sensor OSW Optical Switch module PWM Power Supply module CPU CPU module DMY Dummy cover P W M D M Y C P U O S W D T S D M Y P W M D T...

Page 26: ...is required before temperature distribution measurement 2 Indicated nominal values vary with the refractive index of the optical fiber 3 Sampling range is 2 m for 50 km range selectable from 1 m and...

Page 27: ...nsions Reset Switch Shutdown Switch Serial Port 3 Serial Port 2 Serial Port 1 Network power switch Network interface Operation status LEDs Optical Output LEDs from top to bottom RCV SND LEDs from top...

Page 28: ...k RJ45 Pin No Signal name Function 1 Transmission data 2 Transmission data 3 Receiving data 4 Not used 5 Not used 6 Receiving data 7 Not used 8 Not used LEDs Table Operation Status LED Indicators LED...

Page 29: ...May 30 2022 Switches Table Various Switches Name Purpose Reset Switch Press this switch to reboot the DTSX3000 Shutdown Switch Press this switch to terminate the DTSX3000 safely Network power switch...

Page 30: ...d end non reflection required Optical connector E2000 APC Optical channels 2 channels 4 channels 16 channels Interface Control Controlled by DTSX3000 Display LEDs HRDY RDY Alarm Active channel Power s...

Page 31: ...s Operation status LED LEDs from left HRDY RDY ALARM Channel display LED Channel display LED Channel display LED Operation status LED Operation status LED Optical input Optical output Optical input Op...

Page 32: ...RDY Green Lit when the hardware is operating normally RDY Green Lit when the system is operating normally ALARM Green Flashes when an alarm condition is present Active Channel Green Indicates an activ...

Page 33: ...ule NFCP050 but no other STARDOM module can be mounted on the base module DTSBM10 Dual redundant configuration is not allowed for CPU module and power supply modules DIN rail mounting is not allowed S...

Page 34: ...A1 5 RACK MOUNT KIT The rack mount kit can be used for laying optical fibers in a cabinet Specification Table Specifications of Rack Mount Kit DTRK10 Item Specifications Dimensions W x H x D 482 6 235...

Page 35: ...Sensor for Single Mode Fiber and Optical Switch Module for Single Mode fiber For checking the operation for Single Mode fiber prepare single mode fiber cables on your own Specification Table Specifica...

Page 36: ...he power supply modules see GS 34P02Q13 01E FCN RTU Low Power Autonomous Controller Hardware and GS 34P02Q12 01E FCN Autonomous Controller Hardware A1 8 CPU MODULE This is the CPU module for the auton...

Page 37: ...rial discharge Grounding 100 or less 4 Cooling Natural air cooling 1 Not applicable to DTFB10 Refer to GS 39J06Z40 01EN DTSXL Distributed Temperature Sensor Long Range System for SMF in the case of sp...

Page 38: ...A x F2 x 10 3 9 8 1G where A Total amplitude mm F Frequency Hz The allowable range for total amplitude is shown below Figure Allowable Vibration Range Radio Device Interference to DTSXL System The fol...

Page 39: ...al switch should be installed for maintenance and safety purposes IMPORTANT To conform the power supply system for the DTSXL System to safety standards use a breaker or a switch that conforms to the s...

Page 40: ...line condition Generally the inrush current when the system is comprised of multiple devices will be less than the total sum of the inrush current of each device Output Capacity The output capacity o...

Page 41: ...ure during writing of files to the system memory of the DTSX3000 say during program update may damage the file system Power failure during DTSX3000 operation will not damage the system memory Power fa...

Page 42: ...tem is designed to be cooled by natural air Install a DTSXL System so that the ventilation air flows upward from its bottom to its top as shown below Figure Installing the Base module CAUTION Be sure...

Page 43: ...ck or general purpose control panel install the supplied insulation bushings to insulate the base module from the rack as shown in the diagram Attach the base module to the 19 inch rack cabinet using...

Page 44: ...Each module occupies 2 slots Installation Restrictions and Precautions Power supply module installation restrictions Redundant configuration is not allowed for the power supply module Does not use slo...

Page 45: ...he optical fiber Removal Procedures 1 Remove the optical fiber 2 Remove the screws that hold the module Detach an optical switch module before detaching the DTSX3000 3 Pull the module forward away fro...

Page 46: ...g environmental concerns into consideration Shown below is an example of recommended arrangements Figure Recommended Cabinet or Panel Arrangement Cable duct for input signals DTSXL System Fuse timers...

Page 47: ...ailed fan or cooler will affect the system consider implementing a backup system such as a temperature sensor in the cabinet panel that will report an alarm in the event of a fan or cooler failure Con...

Page 48: ...stem within specified limits by insulating the cabinet or panel from vibration and shock generating sources or fixing the cabinet or panel with rubber isolators For shocks and vibrations generated by...

Page 49: ...Module and Circuit Twist the power supply wires and minimize the wiring length The power supply modules of the DTSXL System incorporate a noise filter and hence have ample immunity to the level of no...

Page 50: ...nication signal wires carry high speed signals Lay these wires in a separate duct at least 10 cm away from other input output wires IMPORTANT Be sure to use specified cables and take extra special car...

Page 51: ...03 m see note below G2 Moderate An environment in which corrosive gas can be detected and it could be determined that the gas is the major cause adversely affecting the reliability of plant equipment...

Page 52: ...00 Cl2 1 2 10 10 NOx 50 125 1250 1250 mm3 m3 Group B HF 1 2 10 10 NH3 500 10000 25000 25000 O3 2 25 100 100 TIP The gas density data indicated in the table are for reference only with the relative hum...

Page 53: ...eous vaporous explosive atmospheres are not generated under normal conditions if generated they exist for a short period of time only CAUTION Control equipment connected to the DTSXL System must not u...

Page 54: ...module 3 DTSX3000 DTSX3000 a0bEcd 1 Distributed Temperature Sensor 4 OSW Module DTOS2L NbEcd 1 2ch Optical Switch DTOS4L NbEcd 1 4ch Optical Switch DTOS16L NbEcd 1 16ch Optical Switch 5 Base Module DT...

Page 55: ...Electrical Code NEC or local electrical codes Other Cables Connect cables according to the National Electrical Code NEC or local electrical codes SEE ALSO For details on the power supply modules NFPW...

Page 56: ...PART A Hardware A 33 IM 39J06B40 01EN 14th Edition May 30 2022 A4 6 2 NFM031 A11 The contents of NFM031 A11 are shown below...

Page 57: ...by type of protection n The DTSXL System is indoor use It can be installed in a hazardous location of Zone 2 after accommodating them in a keyed cabinet 1 of protection rating IP54 or higher It is po...

Page 58: ...module 3 DTSX3000 DTSX3000 a0bEcd 1 Distributed Temperature Sensor 4 OSW Module DTOS2L NbEcd 1 2ch Optical Switch DTOS4L NbEcd 1 4ch Optical Switch DTOS16L NbEcd 1 16ch Optical Switch 5 Base Module DT...

Page 59: ...a metal conduit and wired from a nonhazardous location without applying stress to the end of the cables In addition connect cables according to local electrical codes for explosion proof Ethernet Cab...

Page 60: ...AGISSE D UN EMPLACEMENT NON DANGEREUX Cabinet for DTSXL System DTSXL System must be installed in a lockable metal cabinet with a key Use a cabinet approved by local explosion proof testing organizatio...

Page 61: ...ularly in North America NI devices can be used in Class I Division 2 locations 2 Class I Division 2 This is an location where gaseous vaporous explosive atmospheres are not generated under normal cond...

Page 62: ...U module 3 DTSX3000 DTSX3000 a0bEcd 1 Distributed Temperature Sensor 4 OSW Module DTOS2L NbEcd 1 2ch Optical Switch DTOS4L NbEcd 1 4ch Optical Switch DTOS16L NbEcd 1 16ch Optical Switch 5 Base Module...

Page 63: ...ess to the end of the cables In addition connect cables according to the Canadian Electrical Code CEC or local electrical codes Serial Port of DTSXL System The cable for serial ports of DTSXL System m...

Page 64: ...CT WHILE CIRCUIT IS LIVE UNLESS AREA IS KNOWN TO BE NON HAZARDOUS French AVERTISSEMENT RISQUE D EXPLOSION NEPAS D BRANCHER TANT QUE LE CIRCUIT EST SOUS TENSION MOINS QU IL NE S AGISSE D UN EMPLACEMENT...

Page 65: ...s use crimp on terminal lugs with insulation sleeve Be sure to use ring tongue terminal and crimping tools of the same brand Applicable Cables Insulated cables for industrial equipment such as 600 V p...

Page 66: ...low to connect the power supply cable 1 Detach the protective cover for the terminals on the power supply module as follows Push the locking tab circled area in figure up to unlock it Pull gently to r...

Page 67: ...manual of the power supply module If this connection is not done the power capacity of the NFPW426 may be exceeded and the power supply cut off Do not connect to the analog field power supply termina...

Page 68: ...ted wires IV JIS C3307 Polyvinyl chloride insulated wires for electrical apparatus KIV JIS C3316 600 V grade heat resistant polyvinyl chloride insulated wires HIV JIS C3317 Heatproof vinyl insulated w...

Page 69: ...happens reboot the DTSX3000 or power it on again RS 232C Communication Cable of DTSX3000 Cable Connection If the RS 232 communication ports of the DTSX3000 are used prepare the RJ45 connector cables R...

Page 70: ...terface leave the connector cover for dust protection on to prevent any damage to the connector SEE ALSO When using and performing wiring for the CPU module NFCP050 you should also refer to the STARDO...

Page 71: ...ccording to a user defined measurement sequence Double ended Measurement The figure below shows an example of a connection for double ended measurement In double ended measurement the sensor optical f...

Page 72: ...y result in sensor optical fiber loss bending loss The sensor optical fiber must be free of reflection sensor optical fiber splicing must be by fusion splicing and the far end of the cable must be pro...

Page 73: ...ents and are routed and connected in compliance with cable specifications A5 4 3 Connecting a Sensor optical fiber Follow the procedure below to connect the optical connector of a sensor optical fiber...

Page 74: ...otect it against dust It should be replaced when the sensor optical fiber is disconnected from the DTSX3000 Do not remove its protective cap if the optical output connector on the DTSX3000 is not used...

Page 75: ...00 start measurement and check that distributed temperature measurement is normal If not redo the cable routing and connection If the DTSX3000 still fails to measure temperature distribution normally...

Page 76: ...of the DTSX3000 For the E2000 APC optical connector press down on its lever to unlock it Then hold it with your two fingers and pull it lightly towards you see the Figure below If the optical connect...

Page 77: ...he optical fiber must be connected directly between the optical connector for optical output on the DTSX3000 and the optical connector for optical input on the optical switch module The optical connec...

Page 78: ...vertically into the optical connector of the DTSX3000 or the optical switch module Improper connection lateral movements and insertion by force may damage not only the optical connector but also the...

Page 79: ...optical fibers must cover the intended temperature distribution measurement range Typical optical fibers are resistant up to 80 C If higher temperatures are to be measured heat resistant optical fiber...

Page 80: ...pecial attention when optical fibers must be bent within a tight space If the requirements are not satisfied the optical fibers and optical fibers may exhibit larger loss leading to unstable distribut...

Page 81: ...h the sensor optical fiber connected Temperature calibration must be done when the DTSX3000 is first used or when optical fiber is replaced Moreover re calibration is required if optical fiber charact...

Page 82: ...B i IM 39J06B40 01EN 14th Edition May 30 2022 PART B Software DTSXL Distributed Temperature Sensor Long Range System PART B Software IM 39J06B40 01EN...

Page 83: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...

Page 84: ...odbus communication support allows the DTSX3000 to be controlled from STARDOM or CENTUM File transfer protocols SFTP SCP are supported so measurement data can be retrieved from a host controller or da...

Page 85: ...m minimum difference or slope selected for each section Detection functions Temperature alarm settings Up to 20 alarm sections can be defined Alarm detection sections can be defined with temperature h...

Page 86: ...istration functions Up to 10 users can be registered for use with HTTPS server and SSH server 1 Available when DTSX3000 is used with optical switch DTOS2 DTOS4 or DTOS16 2 The measurement time setting...

Page 87: ...ce operations are allowed when connected with administrator privileges 9 HTTP server 1 The DTSX3000 runs as an HTTP client and transfers measurement data files in LAS format or converted to WITSML for...

Page 88: ...t sends a time request to the SNTP server periodically 2 Smooth time transition is adopted to avoid sudden change in time However immediate time change immediate change to specified time is adopted at...

Page 89: ...ase go to the URL below However the information on the following website does not cover all the possible risks https partner yokogawa com global member ofs index htm B1 3 1 Connecting to Control Devic...

Page 90: ...ent execution and control as well as retrieval and display of measurement data to be done easily using a graphical user interface GUI LAS 2 0 Data Conversion Software included in the DTAP3000 optional...

Page 91: ...browser as needed This section explains how to perform setup for IP address configuration for the DTSX3000 when the previously configured address is unknown or forgotten You need to enter a MAC addre...

Page 92: ...he IP Address Configuration Tool 2 Double click on the Setup exe file in the Configurator IP Address Config Tool folder of the CD ROM On the displayed Welcome window click the Next button 3 Accept the...

Page 93: ...Confirm to begin installation If you have accepted the terms of the license the installation initiation window is displayed Click the Install button to begin installation Installation begins When inst...

Page 94: ...e Should you forget a previously configured IP address put the system in IP address setting selected state when it transits to boot mode selection state immediately after power up If the system is lef...

Page 95: ...01EN 14th Edition May 30 2022 Run the IP Address Configuration Tool Double click on the DTSX IP Address Config Tool icon on Windows desktop The IP Address Configuration Tool runs Configure IP address...

Page 96: ...f the row IP address settings for up to ten DTSX3000 units can be saved 2 Start IP Address Assignment Click the Start button The tool receives the MAC address sent by the DTSX3000 requesting for IP ad...

Page 97: ...nfiguration Tool Click the Stop button to terminate the execution of the IP Address Configuration Tool Perform system configuration After IP Address configuration is completed the DTSX3000 boots up in...

Page 98: ...it transits to boot mode selection state immediately after power up If the system is left in this state for more than 5 seconds it goes into maintenance state The state of the DTSX can be determined f...

Page 99: ...sualization Software DTAP3000 or the DTSX3000 Data Conversion Software WITSML1 3 1 1 DTAP3000D which are available as separate purchases the user must have ReadWrite authority Therefore if you are usi...

Page 100: ...2022 B2 Using the DTSX3000 B2 1 Operation Flowchart The operation flowchart below shows the overall operation flow when using the DTSX3000 for the first time For details on individual items see the r...

Page 101: ...o shutdown the system safely press its SHUTDOWN switch Alternatively you can select Reboot Shutdown from its Web configuration window TIP If you cut off the power supply or reboot the system using the...

Page 102: ...B4 System Operation B4 1 Operation Overview The figure below shows transitions between the various DTSX3000 operation states The current operation state of the DTSX3000 can be determined by checking t...

Page 103: ...in IP address setting state When IP address configuration is completed the DTSX3000 transits to Loading state and then to Maintenance state B4 4 Loading State In Loading state the DTSX3000 loads the...

Page 104: ...9 Test Mode The Test mode is used for fault analysis and system recovery when the system fails to boot The DTSX3000 enters Test mode automatically if the system software is not installed or a fatal er...

Page 105: ...and COM2 communication ports can be used for Modbus communications SEE ALSO For details on Modbus communications see the DTSX3000 Communications Guide IM 39J06B40 02EN For details on Modbus communica...

Page 106: ...etion of generated measurement data files can be executed from a PC connected to the DTSX3000 The following SFTP and SCP client access restrictions apply depending on the user authority level of the l...

Page 107: ...dbus server function You can define the port number and timeout duration for Modbus TCP communications A Modbus client device can connect to the DTSX3000 running as a Modbus server device to read and...

Page 108: ...sages SEE ALSO See Section B6 2 Item List for details TIP For Windows 8 Windows 8 1 or Windows 10 users use the desktop app version of the Internet Explorer as the Web browser When accessing the DTSX3...

Page 109: ...39J06B40 01EN 14th Edition May 30 2022 TIP To modify system configuration data you need to access the DTSX web page using a user ID with Administrator authority If user authentication is successful t...

Page 110: ...ccessing the DTSX3000 from a Web browser The side menu displays a list of items Clicking on an item in the list displays its associated content in the display configuration area The side menu can be s...

Page 111: ...lization Software is not allowed in maintenance state To update or restore Revup Restore DTSX3000 system software you need to additionally acquire and be granted Revup authority after putting the DTSX...

Page 112: ...shown in the figures below Online state Maintenance state No authority granted Maintenance state operation authority granted Maintenance state operation authority granted and system update page displa...

Page 113: ...n May 30 2022 Revup authority granted Revup authority not granted Revup authority granted Button for releasing Revup authority Message indicating that Revup authority is not granted because another us...

Page 114: ...0 2022 B6 1 4 How to Transit between Operation States This subsection describes how to transit between operation states and how to acquire authority Procedure for transiting from online state to maint...

Page 115: ...2022 2 The Reboot page is displayed Select the Reboot Maintenance Mode option and click the OK button 3 When the confirmation window is displayed click the OK button 4 The DTSX3000 reboots Select Rebo...

Page 116: ...y granted TIP Do not reboot the DTSX3000 during temperature measurement Correct measurement result is not guaranteed if reboot is executed during measurement Procedure for transiting from maintenance...

Page 117: ...quiring operation authority in Maintenance no authority granted state 1 Click the acquire auth button 2 Confirm that operation authority has been granted Maintenance state no authority granted Click t...

Page 118: ...indow is displayed As an example click Date after operation authority is granted If another user has acquired authority the DTSX3000 transits to Maintenance no authority granted state without displayi...

Page 119: ...that Revup authority has been granted To release Revup authority click the Unlock button 3 Revup authority is not granted if another user is already granted Revup authority Button for locking Revup au...

Page 120: ...area Alarm display An alarm icon is displayed when any message at or above the configured log alarm level described in Subsection B6 4 16 Log Alarm Level Configuration Log Alarms is logged Notice mes...

Page 121: ...ART B Software B 38 IM 39J06B40 01EN 14th Edition May 30 2022 Error messages Error messages such as a message informing that a specified value is out of the valid data range are displayed Error messag...

Page 122: ...ration restoration Restore system configuration backup data on a PC to DTSX3000 Revup System revision update Perform revision update of the kernel root file system bit data for FPGA and DTSX3000 appli...

Page 123: ...O Informations RAS O O O O O Sys log O O O O O App log O O O O O Configurations User O X O O O X O X O X Network O X O O O X O X O X Date O X O O O X O X O X Web O X O O O X O X O X SSH SFTP O X O O O...

Page 124: ...side menu Ten historical RAS information records from ras0 to ras9 can be displayed with ras0 representing the most recent information and ras9 representing the oldest information The oldest RAS infor...

Page 125: ...y applications application log in log display by selecting Sys log or App log from the side menu Log messages at or above the log alarm level described in Subsection B6 4 16 Log Alarm Level Configurat...

Page 126: ...ment User You can add delete or change the password of a DTSX3000 user in user management Change password Select a user type from Admin User ReadOnly User ReadWrite List of registered users Delete a u...

Page 127: ...nd be up to 31 characters long Role Select the user type from among Admin User readwrite and User readonly Password Specify a password A password must begin with an alphanumeric character and consist...

Page 128: ...PART B Software B 45 IM 39J06B40 01EN 14th Edition May 30 2022 Tip The dtsx user cannot be deleted Deleting a user does not remove the user s home directory You cannot delete the root user...

Page 129: ...other network related settings in network configuration IP address settings for DNS Domain name settings Enable or disable security function for DTSX3000 communications Hostname IP address subnet mas...

Page 130: ...longer than 15 characters Domain0 Domain1 Up to 2 domain names can be specified A domain name must begin and end with an alphanumeric character consist only of alphanumeric characters the period char...

Page 131: ...equired settings Date can be specified by directly entering a date in mm dd yy format in the Date field or by selecting a date from the displayed calendar Time can be specified directly by entering a...

Page 132: ...01EN 14th Edition May 30 2022 Specify a date from 01 01 2001 to 12 31 2037 Display input helper window Specify a time from 00 00 00 to 23 59 59 Please select Settngs with selected checkboxes If settin...

Page 133: ...zone can be specified using the following format Keywords enclosed within square parentheses are optional std and dst Time zone abbreviations comprising 3 or more letters std represents standard time...

Page 134: ...to 365 2 29 if present is included in the calculation Mm n d month week and day M is a prefix m is an integer from 1 to 12 representing a month n is an integer from 1 to 5 representing the nth week of...

Page 135: ...required settings press the Convert button A time zone string is converted automatically from the specified settings Click the Copy button to copy the automatically generated character string to the...

Page 136: ...g time zone with the time zone setting helper window using the Week setting method Select week for the Setting method After specifying the required settings press the Convert button Select the Week se...

Page 137: ...oftware B 54 IM 39J06B40 01EN 14th Edition May 30 2022 A time zone string is converted automatically from the specified settings Converted timezone string Copy converted string to Timezone field in Da...

Page 138: ...can access the DTSX3000 system settings Non Authentication Top Dir0 Non Authentication Top Dir1 These settings are for maintenance purpose and are normally not used HTTPS Enable Specify whether to ena...

Page 139: ...ect no for HTTPS Enable 3 Reboot the DTSX3000 in maintenance state 4 Perform the steps described for Enabling HTTPS above Tip An SSL server certificate must be addressed to the hostname domain name of...

Page 140: ...e SSH Be sure to select yes if connecting to the DTSX3000 using the DTSX3000 Control Visualization Software or the DTSX3000 Data Conversion Software TELNET If using TELNET select yes if not using TELN...

Page 141: ...M2 Speed Baud rate bps 1 2 Data length Data length 7 or 8 3 7 or 8 3 Stop bit Number of stop bits 1 3 or 8 1 3 or 2 Parity Parity check None 3 even or odd None 3 eve or odd Flow control Flow control N...

Page 142: ...B Software B 59 IM 39J06B40 01EN 14th Edition May 30 2022 B6 4 9 PPP Server Configuration PPP This configuration item is intended for maintenance use and is normally not used Set PPP Enable to no Sele...

Page 143: ...haracters An IP address must consist only of numeric characters and the period character and be no longer than 15 characters After specifying the required settings click the Set button The DTSX3000 co...

Page 144: ...0 configuration is updated but the changes are applied only after reboot B6 4 12 System Configuration Backup Backup You can back up all system configuration data to a PC in system configuration backup...

Page 145: ...ore DTSX3000 system configuration data from a configuration archive file saved previously on a PC using system configuration backup Press the Browse button On the displayed file selection dialog selec...

Page 146: ...and DTSX3000 applications in system revision update by selecting Revup from the side menu SEE ALSO See Section B11 6 System Revision Update for details B6 4 15 Save Log Saveinfo You can download RAS...

Page 147: ...ied level is logged an alarm is displayed as shown in Subsection B6 1 5 Auxiliary Information Display and the log message will be displayed as highlighted as shown in Subsection B6 4 2 Log Display Sys...

Page 148: ...utdown the DTSX3000 as well as select the boot mode operation state of the DTSX3000 after reboot in reboot shutdown by selecting Reboot from the side menu Boot up in online mode Boot up in IP address...

Page 149: ...system goes into power save mode The system exits from power save mode when it receives a start measurement instruction or start self test instruction The system goes into power save mode immediately...

Page 150: ...dbus TCP Available port numbers are 502 and from 1024 to 65535 Default value is 502 Port Number Specify the port number to be used by Modbus TCP Available port numbers are 502 and from 1024 to 65535 D...

Page 151: ...or to save files without compression If you have selected to save files in compressed format files will be stored in gzip compression format with file extension gz To specify the converted file forma...

Page 152: ...FAST TOOLS needs to delete files in the mnt data LATEST directory so it is necessary to grant write access to the mnt data LATEST directory If you are not using FAST TOOLS be sure to set Write Access...

Page 153: ...sferred files Check the Delete files selection and then click the Set button The following confirmation window is displayed Check the settings and if it is alright to delete the files click OK Tip Del...

Page 154: ...l shorten the measurement time by omitting the creation of data files used for communication between DTSX3000 and the software The default value is valid Tip The measurement time saving achievable by...

Page 155: ...assword must be at least 6 characters long Error Passwords don t match The entered password and confirmation password do not match Error user exists username A specified username is already registered...

Page 156: ...ow Message Description Error Local IP address and remote IP address is a duplicate value The specified local IP address and remote IP address are the same Error Passwords are less than six characters...

Page 157: ...cvsoft The specified port number conflicts with the port number used for cvsoft Modbus serial related settings window Message Description Error Slave address that can be used are from 1 to 247 The spe...

Page 158: ...the execution order within one measurement sequence using a sequence table as described later in Subsection B7 2 1 Sequence Table Settings and then define the execution order between measurement sequ...

Page 159: ...ble it Principle Select Single End for single ended measurement or Double End for double ended measurement Channel Near Select a channel from Ch1 to Ch16 as the forward direction channel in double end...

Page 160: ...d measurement interval the next measurement begins immediately after the current measurement ends SEE ALSO For details on double ended measurement see Subsection B1 1 2 Double ended Measurement Settin...

Page 161: ...6B40 01EN 14th Edition May 30 2022 Single Continuous Sequence Interval Sequence Start Time Every Day One Cycle Sequence Mode Repetition Mode Start Time Setting Specify up to 6 values in Start Time Tab...

Page 162: ...l between two start times in the sequence table as a value from 0 seconds to 86400 seconds in units of 1 second The sample settings shown in the preceding screen capture defines the following sequence...

Page 163: ...n above where Start Time Setting is 1 Cycle the measurement sequence is executed thrice at 00 00 07 00 and 14 00 hours respectively after which the system goes into idle mode sec 1st sequence executio...

Page 164: ...4th Edition May 30 2022 In the example shown above the measurement sequence is executed every day at 01 00 hours sec 2nd sequence execution Day 1 01 00 hours Day 2 01 00 hours Day 3 01 00 hours 3rd se...

Page 165: ...arger measurement error due to reflection if the optical fiber used does not have anti reflection termination so standard rate should be selected unless there is some special reason not to do so Dista...

Page 166: ...The measurement time excludes alarm detection time sectional data generation time and measurement data file creation transfer time Tip The Averaging parameter determines the number of measurements use...

Page 167: ...the far end of the sensor fiber falls below the specified Threshold detection threshold value during measurement Specify a threshold level from 99 9 dB to 0 0 dB in units of 0 1 dB If On is not select...

Page 168: ...d for calculation from among Total Length All Data and Range Setting The distance axis range specified here is used for temperature data output Modbus or measurement data files alarm detection and sec...

Page 169: ...up Index Specify the group refractive index of a fiber section as a value from 1 3 to 1 79999 in units of 0 00001 Distance is calculated using the specified group refractive index Loss Specify the tra...

Page 170: ...ected If fiber failure happens the measurement results up to the fiber failure position of the forward and reverse directions are combined Winding Coeff Specify the winding coefficient of a fiber sect...

Page 171: ...alculating the total differential loss between the Stokes and anti Stokes light over the full fiber length You can select either to use a known differential loss value or use the execution result of d...

Page 172: ...reverse distance from the sensor optical fiber end E Distance Of E To P2 Distance Of W To P2 To specify the far end P2 as a forward distance from the switch position W select Distance Of W To P2 and s...

Page 173: ...the position for temperature calibration and the temperature averaging distance of the DTSX3000 Position Specify the position for temperature calibration as a value from 0 0 m to 50000 0 m in units o...

Page 174: ...Specify the temperature offset as a value from 273 15 deg C to 999 99 degC in units of 0 01 deg C The corrected measured temperature is calculated using the following formula Corrected temperature C L...

Page 175: ...purchase the DTSX3000 allows up to 16 optical fibers to be connected for multi channel measurement by auto channel switching The optical switch is available in 3 models supporting 2 channels 4 channe...

Page 176: ...transferred to an external HTTP server using the HTTP transmission function of the DTSX3000 Data Output Conditions The DTSX3000 can generate LAS formatted and CSV formatted converted data files The da...

Page 177: ...OUNTRY Country name 7 SERVICE Corp Service company name 8 DATE Measurement start date and time Specify the output format for measurement start date and time and whether to use UTC or local time repres...

Page 178: ...ion and interval between data in the depth direction STEP indicates the depth interval between measured data if the interval is constant and indicates 0 if the interval is variable A constant value is...

Page 179: ...p 1 COMPANY Company name 2 WELL Well name 3 FIELD Field name 4 LOCATION Location name 5 PROVINCE Province name 6 COUNTY County name STATE State name COUNTRY Country name 7 SERVICE Corp Service company...

Page 180: ...number of data values contained in the data section varies with the Calculation Range setting SEE ALSO For details on the calculation range see Subsection B7 3 1 Fiber Settings Well info section Data...

Page 181: ...at string other than the format specifiers are output without conversion Format Specifiers Format specifier Data Example Remarks Y Year in Gregorian calendar 2011 Four digit representation y 11 Two di...

Page 182: ...You can select whether to have the DTSX3000 compress converted files before archival For more details see Subsection B6 5 4 Data Conversion Related Settings The DTSX3000 assigns a user specified strin...

Page 183: ...specified format string other than the format specifiers are output without conversion Format specifiers Format specifier Data Example Remarks Y Year in Gregorian calendar 2011 Four digit representat...

Page 184: ...ing counter from 000000 to 999999 assigned to each created file The counter returns to 000000 when it reaches 999999 mm 00 to 99 nn 00 to 99 pp 00 to 99 mm and nn together refers to a directory where...

Page 185: ...cluded in the auto assigned part of the file name of a file stored thereunder The figure below shows an example of a LAS archive directory The mnt data LATEST directory stores up to 1200 links to the...

Page 186: ...ent The procedures described in this subsection assume the use of SFTP client software WinSCP Ver4 3 3 Getting individual files from the mnt data BACKUP directory 1 Connect to DTSX3000 Run WinSCP Ente...

Page 187: ...SH SFTP client may display a host key does not match message as shown below If this message is displayed there is a possibility that you have connected to the computer of a malicious attacker pretendi...

Page 188: ...e current local directory and the current remote directory 3 Get files Select one or more files to be retrieved from the remote directory side and then drag and drop it into the local directory side C...

Page 189: ...name already exists in the local directory side before transfer a confirmation dialog is displayed By clicking No to All you can shorten transfer time by retrieving only files that are not already pr...

Page 190: ...Synchronization options as shown in the dialog below and press the OK button When retrieval is completed all files not present in the local directory are retrieved Select file existence as the only sy...

Page 191: ...urement new files are added constantly to the mnt data LATEST directory Repeat steps 2 and 3 to perform resynchronization Only files newly added to the mnt data LATEST directory will be retrieved to t...

Page 192: ...via Modbus or DTSX3000 Data Conversion Software SEE ALSO For details see the DTAP3000 LAS2 0 Guide IM39J02B40 02EN For details on configuration via Modbus see the DTSX3000 Communications Guide IM39J06...

Page 193: ...e xxxxxx is the serial number of the file which failed to be transmitted If one of the following occurs except for the above condition transmission will start with the file which is midway between the...

Page 194: ...ion directory for storing received files and specifies whether user authentication is required Defines the destination directory var www html webdav if webdav is specified in the URI Basic authenticat...

Page 195: ...d password appropriate for the desired destination directory Specify a URI Destination directory on HTTP server side https hostname or IP address webdav var www html webdav https hostname or IP addres...

Page 196: ...transmission method install the CGI program itself in a directory requiring authentication on the server side You can require or skip authentication by specifying the corresponding CGI program in the...

Page 197: ...ware B 114 IM 39J06B40 01EN 14th Edition May 30 2022 CGI programs user authentication is not required Form data name Specify this in server transmit file settings Destination directory for storing rec...

Page 198: ...wn below Transmission settings Specify a CGI file form name uploadfile in this example user ID and password appropriate for the desired destination directory Specify a URI CGI file Destination directo...

Page 199: ...EN To use the WITSML data conversion function however you need to first install its license using the DTSX3000 Data Conversion Software WITSML1 3 1 1 The procedures for installing and deleting a licen...

Page 200: ...nstallation is successful the following dialog is displayed Press the OK button Next reboot the DTSX3000 This completes license installation 5 Reboot DTSX3000 Terminate and restart the DTSX3000 Data C...

Page 201: ...TSML1 3 1 1 Next connect to the DTSX3000 SEE ALSO For details on connection configuration see the DTAP3000D Guide IM39J02B40 03EN 2 Open the help window Press the icon displayed at the top right of th...

Page 202: ...14th Edition May 30 2022 To delete the license press the OK button 4 Complete license deletion If license deletion is successful the following dialog is displayed Press the OK button Next reboot the D...

Page 203: ...and output of the result of alarm detection is described in Section B9 3 Alarm Results B9 2 Configuration of Alarm Conditions Alarm conditions can be configured via Modbus SEE ALSO For details see th...

Page 204: ...s calculated automatically and thus need not be specified When specifying contiguous alarm sections where the Start Depth of a section coincides with the End Depth of the preceding section you can sav...

Page 205: ...arm when any measured value within the section satisfies the alarm condition Select Average to report an alarm when the average measured value of a section satisfies the alarm condition Temperature Hi...

Page 206: ...y Temperature High Limit Specify a delay temperature high limit value from 273 15 to 999 99 An alarm is reported when the measured value exceeds the limit for a specified number of successive measurem...

Page 207: ...range see Subsection B7 3 1 Fiber Settings B9 3 Alarm Results As described in Section B9 2 Configuration of Alarm Conditions alarm conditions can be configured for up to 50 alarm sections Alarm detect...

Page 208: ...cted alarm The table below describes what is displayed for the alarm position Depth when one or more alarms have been detected for a channel Alarm Type Depth Average type alarm detected The Start Dept...

Page 209: ...onfiguration of Sectional Data Conditions and the output result of sectional data generation is described in Section B10 3 Sectional Data Results B10 2 Configuration of Sectional Data Conditions Secti...

Page 210: ...retrievable for each unit If the Depth of a section is not divisible by the specified Unit the last unit will be shorter than the specified value For example if Depth 100 m and Unit 30 m then the sec...

Page 211: ...e generated for up to 200 sections The figure below shows the relationship between configuration sections and result sections using the example given in Section B10 2 Configuration of Sectional Data C...

Page 212: ...s the Web browser B11 1 1 Access Procedure To access the DTSX3000 enter the following URL in the Web browser Tip If HTTPS communication is enabled specify https A user authentication dialog is display...

Page 213: ...Software B 130 IM 39J06B40 01EN 14th Edition May 30 2022 SEE ALSO For details on user authentication as well as enabling and disabling of HTTPS communication see Subsection B6 4 6 Web Server Configura...

Page 214: ...the system is stopped The oldest RAS information is deleted and new RAS information is added each time the system is started ras0 represents the most recent information while ras9 represents the oldes...

Page 215: ...PART B Software B 132 IM 39J06B40 01EN 14th Edition May 30 2022...

Page 216: ...king Alarms Alarm display indicates whether an alarm has been detected There are two types of message logs related to alarms namely system log and application log The former is output by the boot load...

Page 217: ...File for details When a message at or above the log alarm level is output to the application log An AppLogAlarm icon is displayed You can check the details of the application log by selecting Informa...

Page 218: ...ecution see the DTSX3000 Communications Guide IM39J06B40 02EN Tip Self diagnosis cannot be executed during measurement If a self diagnosis execution command is received during measurement an execution...

Page 219: ...ias Voltage ST NG data APD Bias Voltage AS NG data Photoreceiver amplifier offset voltage Amp Offset Voltage ST OK data Amp Offset Voltage AS OK data Amp Offset Voltage ST NG data Amp Offset Voltage A...

Page 220: ...DTSX3000 from a Web browser you need to enter a pre defined user name and password SEE ALSO For details on the access procedure see Subsection B11 1 1 Access Procedure If access is allowed the Top pa...

Page 221: ...lecting the type of log file to be saved Select the System Log option and click the Download button On the displayed window click the Save button After specifying an appropriate folder on the Windows...

Page 222: ...PART B Software B 139 IM 39J06B40 01EN 14th Edition May 30 2022...

Page 223: ...Subsection B6 4 16 Log Alarm Level Configuration Log Alarms are highlighted in red in the display If any log message at or above the specified Applicationlog Alarm Level is present an AppLogAlarm icon...

Page 224: ...ion and click the Download button On the displayed window click the Save button After specifying an appropriate folder on the Windows PC and an output file name click the Save button The log file is s...

Page 225: ...update after purchase please use the checksum file that is distributed with the archive container file SEE ALSO For details on performing revision update see Subsection B11 6 System Revision Update 1...

Page 226: ...ou need to perform tamper check for system files The tamper check for system files checks the system image as well as files that are required for system operation and are not to be converted The follo...

Page 227: ...ult against the content of the uImage R checksum txt file R denotes the version number For instance if the version number is R1 02 01 the output result should be compared against the uImage R1 02 01 c...

Page 228: ...ent is in progress or when the DTSX3000 is in maintenance state Checking loaded system files Phase 1 Checking base executable files OK Phase 2 Checking initialization script files on boot sequence OK...

Page 229: ...e figure below shows the content of a sample app R checksum txt file Check that the output result obtained above is identical to line 4 of the app R checksum txt file Compare the output result of phas...

Page 230: ...on cycle weekly var log lighttpd error log var log lighrrpd access log Checking cycle for rotation hourly Number of historical log files 1 Size threshold for rotation 6 MB Rotation cycle weekly In the...

Page 231: ...ogin request from the SSH SFTP the client receives the following message and connection fails Received message too long 1416589088 If this message is displayed check the number of existing connections...

Page 232: ...en done and if so update the SSH SFTP client with the new host key sent by the DTSX3000 2 Move to the directory containing the log file Change the current remote directory source for copying and the c...

Page 233: ...PART B Software B 150 IM 39J06B40 01EN 14th Edition May 30 2022...

Page 234: ...created configuration archive file later B11 5 1 Backup To perform backup follow the procedure below 1 Boot up the DTSX3000 in online state or maintenance state SEE ALSO For details on the boot proced...

Page 235: ...22 3 Display the backup window If the system was started in online state select Tool Backup from the side menu If the system was started in maintenance state click the acquire auth button at the top o...

Page 236: ...ment click the Execute Backup button Tip Comment is optional 5 Specify the destination folder for saving the configuration archive file When you click the Save button a destination folder selection wi...

Page 237: ...used as the configuration archive file for restoration However restoration is not guaranteed to work properly if the configuration archive file has been edited using a PC or some other means after ba...

Page 238: ...PART B Software B 155 IM 39J06B40 01EN 14th Edition May 30 2022 3 Acquire authority Click the acquire auth button at the top of the Top page The Tool menu is displayed in the side menu...

Page 239: ...PART B Software B 156 IM 39J06B40 01EN 14th Edition May 30 2022...

Page 240: ...ther operations To unlock is only possible in the locked browser If you close the browser in a locked status lock status will continue for 10 minutes During the period you and the other users can not...

Page 241: ...PART B Software B 158 IM 39J06B40 01EN 14th Edition May 30 2022 If restoration is successful the following window is displayed...

Page 242: ...PART B Software B 159 IM 39J06B40 01EN 14th Edition May 30 2022 7 To unlock the other operations Push Unlock button to unlock the other operations...

Page 243: ...p File System An archive container file is a collection of their respective revision update images within one file The latest archive container file can be downloaded from the members only Yokogawa Pl...

Page 244: ...PART B Software B 161 IM 39J06B40 01EN 14th Edition May 30 2022 4 Aquire authority Click the acquire auth button at the top of the Top page The Tool menu is displayed in the side menu...

Page 245: ...J06B40 01EN 14th Edition May 30 2022 5 Transit to Revision Update mode and get Revision Update Revup authority Select Tool Revup from the side menu On the displayed Revision Update window click the Lo...

Page 246: ...lick the Browse button on the Container row and select the archive container file On the System Configuration row select Keep to keep the system configuration file or select Overwrite to overwrite it...

Page 247: ...PART B Software B 164 IM 39J06B40 01EN 14th Edition May 30 2022...

Page 248: ...update Click the Begin button at the bottom of the window displayed when uploading is completed Tip The FPGA Config is not included in the archive container file Revision update of the FPGA Config is...

Page 249: ...PART B Software B 166 IM 39J06B40 01EN 14th Edition May 30 2022 When revision update is completed the following window is displayed...

Page 250: ...enerated on the DTSX3000 This may cause a host name does not match error message to be displayed on the first SSH SFTP connection to the DTSX3000 after the system update If this happens check with the...

Page 251: ...ome pre setup which is described in the subsequent subsections B12 2 Retrieving Data Using FTP SFTP To retrieve temperature data WITSML files from DTSX3000 using the FTP or SFTP client function of FAS...

Page 252: ...C i IM 39J06B40 01EN 14th Edition May 30 2022 PART C Operation and Maintenance DTSXL Distributed Temperature Sensor Long Range System PART C Operation and Maintenance IM 39J06B40 01EN...

Page 253: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...

Page 254: ...ns its previous idle state C1 1 2 Power off Procedure To power off the DTSX3000 safely you need to put it into shutdown state ready for power down state before turning off the power If a CPU module is...

Page 255: ...M 39J06B40 01EN 14th Edition May 30 2022 CAUTION The RDY LED blinks during SHUTDOWN processing Do not switch off the power supply unplug the power cord or press the RESET switch until SHUTDOWN process...

Page 256: ...he CPU module C1 3 System Operation upon Power Failure If the power supply to the system is interrupted for a period exceeding the instantaneous power failure sense delay time stipulated in the specif...

Page 257: ...nt Single ended measurement can be used in distributed temperature measurement applications where the sensor optical fiber cannot be looped back or no optical switch is used Single ended measurement i...

Page 258: ...th Edition May 30 2022 Figure Differential Loss Correction Signal level dB Distance m Anti Stokes pre correction Stokes Anti Stokes post correction Temperature degC Distance m Post correction Pre corr...

Page 259: ...Stokes light over the full sensor optical fiber length is still required to correct for absolute temperature error in double ended measurement Two options are provided for calculating this differenti...

Page 260: ...ical fiber failure is detected at any position on the sensor optical fiber measurement results of the forward and reverse directions up to the sensor optical fiber failure position are combined Signal...

Page 261: ...ition E denotes the sensor optical fiber end position Figure Sensor optical fiber Failure Illustration Diagram Signal level dB Anti Stokes Reverse Stokes Forward Stokes Reverse Anti Stokes Forward W E...

Page 262: ...ment resolution depends on sensor optical fiber loss and is thus comparatively better near the DTSX3000 and worse at the far end On the contrary in double ended measurement temperature measurement res...

Page 263: ...Section B7 3 3 Calibration C2 2 1 Temperature Calibration Procedure To perform temperature calibration follow the steps described below 1 Determine sensor optical fiber sections Temperature calibratio...

Page 264: ...the same lot from the same manufacturer the same value determined for any one section can be used for other sections 4 Calibration Finally you need to configure parameters T1 T1 for single point temp...

Page 265: ...urement time required for sampling resolutions of 50 cm are 2 times respectively at the same temperature measurement accuracy C3 1 1 Effect of Spatial Resolution The DTSX3000 has a spatial resolution...

Page 266: ...d sensor optical fiber ends may be incorrect due to the effect of Rayleigh scattering C3 1 3 Effect of Signal Processing The DTSXL uses digital signal processing which improves the temperature resolut...

Page 267: ...the measured values vary within a range 3 of the median temperature Quadrupling the measurement time halves the temperature resolution The absolute temperature measurement accuracy depends on the tem...

Page 268: ...ris Installation Is each module installed properly No looseness Are the connectors for cables inserted properly and securely attached No looseness Are any of the external wiring screws loose No loosen...

Page 269: ...ing maintenance use a grounded wrist strap with a 1M grounding resistance Be sure to ground the wrist strap When performing operations on a table be sure to perform operation on an anti static sheet t...

Page 270: ...TDOWN switch If a CPU module is installed shut down the CPU in accordance with its replacement procedure 2 Turn off the power 3 Remove the sensor optical fiber Replace the dust protective cap of the o...

Page 271: ...cordance with its replacement procedure Replacing CPU Module Follow the procedure below to replace a CPU module 1 Save retentive data to the on board Flash Memory 2 Backup system files using the backu...

Page 272: ...ery Product name Part number Battery S9883FA Recommended Replacement Period The battery lifespan greatly depends on ambient temperature The following table shows the ambient temperatures and recommend...

Page 273: ...e the STARDOM FCN FCJ Guide IM 34P02Q01 01E Replacing the Battery of DTSX3000 Preparation Have on hand a new battery as well as the necessary equipment for connecting to the DTSX3000 from a PC Replace...

Page 274: ...J06B40 01EN 14th Edition May 30 2022 3 Pull out the upper part of the battery followed by its lower part 1 Pull out the upper part of the battery 2 Remove the whole battery Figure Removing the Battery...

Page 275: ...SET switch When you press the RESET switch the module boots up in Maintenance mode with blinking HRDY and RDY LED indicators 8 Display the DTSX3000 maintenance menu from a Web browser running on a PC...

Page 276: ...s charged the electrolyte inside the battery may bubble or the internal pressure may rise due to generated gas When this happens the battery may leak electrolyte generate heat burst or ignite Do not u...

Page 277: ...and leak electrolyte generate heat burst or ignite Do not drop or throw the battery or otherwise apply strong impact The battery may leak electrolyte generate heat burst or ignite Do not deform the ba...

Page 278: ...his directive is only valid in the EU Batteries are included in this product When you remove batteries from this product and dispose them discard them in accordance with domestic law concerning dispos...

Page 279: ...Data Size 1 Measurement Repetition Period 16000 1 minute 6 years 16000 10 minutes No replacement required 1 The data size is the distance range divided by the sampling resolution For example assuming...

Page 280: ...tly using your fingers Install the new SD card Ensure that write protection of the SD card is switched off i e the SD card is writeable Insert the SD card with correct orientation and press it in gent...

Page 281: ...ack of the DTSX3000 Optical Switch Modules Optical switch modules may require regular replacement depending on their usage and installation environment Table Recommended Replacement Period Operating C...

Page 282: ...sensor optical fiber fails or degrades measurement may fail completely or the rated performance is not achieved A defective sensor optical fiber must be repaired or replaced After repairing or replaci...

Page 283: ...rature humidity vibration corrosive gases ventilation external noise etc meet the installation specification of the DTSX3000 SEE ALSO For details on the installation specification see Chapter A2 Insta...

Page 284: ...rovides a self diagnosis function for detecting hardware problems Some of the tests of self diagnosis are automatically executed by the DTSX3000 at system startup while waiting for measurement and at...

Page 285: ...nector Replace the sensor optical fiber with another optical fiber or the Optical Fiber for DTSX3000 DTFB10 and check if measurement can be correctly executed If there is no problem with the optical f...

Page 286: ...ing the communication cable properly Try replacing the communications cable with a new cable Check if firewall and other settings of communication devices are proper Check if IP addresses and subnet m...

Page 287: ...Edition R1 03 01 or later Update of compliant standards for single mode fiber May 2019 9th Edition R1 03 01 or later Update of compliant standards and changing of related description Symbol mark and...

Page 288: ...ition May 30 2022 Contact information Phone 81 422 52 5735 E Mail OFS_Inquiry cs jp yokogawa com Author Yokogawa Electric Corporation Publisher Yokogawa Electric Corporation 2 9 32 Nakacho Musashino s...

Page 289: ...IM 39J06B40 01EN 14th Edition May 30 2022 Blank Page...