DIQ/S 28X
Outputs
141
ba77168d02
03/2017
In order to set up the
Frequency controller
function for a relay output, the relay
output must be linked with a sensor (see section 7.4).
Settings
Characteristic
curve
If a value for
End value
that is greater than the
Start value
is entered, the output
has a positive characteristic curve.
In order to obtain a negative characteristic curve, a value for
End value
must be
entered that is smaller than the value for
Start value
.
Setting
Selection/Values
Explanation
Measured variable
Main variable
Adjoining variable
Main variable
designates the
actual measured parameter
of the sensor (e.g. pH, oxy-
gen, etc.).
Adjoining variable
desig-
nates an additional mea-
sured parameter (e.g.
temperature).
Start value
within the measur-
ing range (sensor-
dependent)
Minimum spacing:
5 % of the measuring range
End value
Frequency (f) min.
0 to 120 1/min
Minimum spacing:
10 1/min
Frequency (f) max.
Frequency with error
0 to 120 1/min
In case of system errors or
sensor errors (see
page 156), the relay
switches with the frequency
specified.
Action
Open
Close
Relay action (see
section 7.6.1)