Ethernet connection (variant -E, -EF)
DIQ/S 28X
116
ba77168d02
03/2017
6.2
Ethernet connection if mounted at an open air test site (DIQ/
S 28X-E[F])
The RJ45 socket does not protect sufficiently against moisture. At an open test
site, the moisture protection device (ADA/E) available as an accessory has to be
mounted for safe Ethernet communication (see section 11.
6.3
Establishing the connection with the DIQ/S 28X
via a network
6.3.1 Opening the
IQ W
EB
C
ONNECT
The web server of the
DIQ/S 28X
provides functions
for the (remote) operation
of the DIQ/S 28X
and for the data exchange (e.g. with a PC) via a network
address.
Prerequisites
All network components (Universal Transmitter, router, network-capable
devices with browser)
are connected
The network services are active
Web browser (HTML 5)
Procedure
Enter the network address of the DIQ/S 28X in the address line in the web
browser
The IP address and the other network settings for the Universal
Transmitter DIQ/S 28X are configured in the
System settings ->
TCP/IP settings
menu.
1
Enter the network address of the
Web browser
DIQ/S 28X-E[F] in the
address line of the web browser.
in the LAN
e.g. name or IP address of the DIQ/S 28X-E[F]
on the Internet,
e.g. fixed IP address of the router or DynDNS name
The network connection to the DIQ/S 28X-E[F] is established.
The start page of the IQ S
ENSOR
N
ET
is displayed.