DIQ/S 28X
Settings/setup
109
ba77168d02
03/2017
5.10.1 Setting the recording interval (
dt
) and recording duration (
Dur.
)
Fig. 5-20 Measured value logging
15 min
0 d
15 d
30 d
45 d
...
5400 d
30 min
0 d
30 d
60 d
90 d
...
10800 d
60 min
0 d
60 d
120 d
180 d
...
21600 d
Recording interval
Possible recording durations with the given recording interval
1
Open the
Einstellungen/Settings
menu with
<S>
.
2
Using
<
>
and
<OK>
, select and confirm the menu item,
Mea-
sured value logging
.
The
Measured value logging
display opens.
It contains a list of all sensors and inactive sensors.
3
Press
<
>
and
<OK>
to select and confirm the
dt
column.
4
Press
<
>
and
<OK>
to select and confirm a sensor.
5
Determine the recording interval with
<
>
and
<OK>
.
When the recording interval is changed, the recording duration changes
at the same time (
Dur.
), because the recording duration results from the
recording interval due to a fixed number of memory blocks.
6
Switch to the selection of columns with
<ESC>
.
7
Press
<
>
and
<OK>
to select and confirm the
Dur.
column.
8
Press
<
>
and
<OK>
to select and confirm a sensor.