DIQ/S 28X
What to do if ...
163
ba77168d02
03/2017
Operating
sequence in case 2
Fig. 9-1
510 - Add/replace sensor
Case 2:
The type of sensor is identical with
the type of sensor in an inactive
dataset (or several inactive datas-
ets), but the serial numbers differ.
Operator intervention is required here.
The connected IQ sensor can:
– be assigned to the inactive dataset
(or one of the inactive datasets).
Example: Replacing an IQ sensor.
Make sure that the substitute sen-
sor has at least the software status
of the active IQ sensor.
– be included as a new module in the
list of the sensors.
The operating sequence to do this is
described below.
1
Connect a new IQ sensor.
2
Switch to the measured value display with
<M>
. The component data-
base is updated. The following display appears (example):
3
Select the required option with
<
>
and confirm with
<OK>
.
– If
Add new sensor
was selected, the system changes directly to the
measured value display. As soon as the IQ sensor is ready for oper-
ation, it delivers a measured value.
– If
Assign sensor as a substitute
was selected, a list of matching inac-
tive datasets appears: