Settings/setup
DIQ/S 28X
100
ba77168d02
03/2017
5.8.3 Alarm output to display
When an alarm event occurs, a window with a text message appears.
Limit value
Lower limit
Upper limit
Type of the alarm event.
Lower limit
: An alarm is triggered if the
defined limit value is undercut.
Upper limit
: An alarm is triggered if the
defined limit value is exceeded.
Lower limit
/
Upper limit
within the measuring
range (sensor-depen-
dent)
Limit value for the alarm event
Hysteresis
0 - 10 % of the measur-
ing range
Hysteresis for the limit value
Designation
(max. 20 characters)
User-defined designation for easier identi-
fication in the alarm message.
Relay output
Dxx / .../Ry
...
...
...
No relay output
Opens a list with all relay outputs where
the
Alarm contact
is set.
Dxx: number of the output module
.../Ry: relay output channel
Here you can select a relay output. When
the alarm event occurs, it carries out the
specified action (Open or Close). For
details, see section 5.8.3 A
Accept
The settings are taken over in the setting
table by pressing
<OK>
. The display
switches to the next higher level.
Cancel
The display switches to the next higher
level without storing the new settings.
Menu item
Selection/Values
Explanations