DIQ/S 28X
Technical data
173
ba77168d02
03/2017
Connecting
terminals
Cable glands
EMC product and
system character-
istics
Max. initial output volt-
age
13 V, in the case of missing or incorrect burden
Accuracy
0.3 % of current value ± 50 µA,
load max. 500
Ω
Functions
Programmable:
– Recorder attenuation, adjustable
0 - 40 mA/s
– Error behavior can be set as required
0 ... 21 mA
– Error behavior according to Namur NE43 can be
selected
– Positive and negative characteristic curve
– PID controller
Terminal type
Screw-type terminal strip, accessible by raising the
lid
Terminal ranges
Solid wires:
Flexible wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
0.2 ... 2.5 mm
2
Suitable for cable diam-
eter
4.5 ... 10 mm or 7 ... 13 mm
EN 61326
EMC requirements for electrical resources for con-
trol technology and laboratory use
– Resources for industrial areas, intended for indis-
pensable operation
– Interference emission limits for resources of class
A
System lightning pro-
tection
Extended protective characteristics as opposed to
EN 61326
FCC, class A
Any combination of the IQ S
ENSOR
N
ET
with IQ S
ENSOR
N
ET
prod-
ucts in a user-specific system achieves the listed EMC characteris-
tics.