Settings/setup
DIQ/S 28X
92
ba77168d02
03/2017
correction function (e.g. FDO
®
700 IQ).
– The automatic (dynamic) correction is enabled in the setting menu for the
sensor with correction function (e.g. FDO
®
700 IQ).
– A sensor measures the influence quantity and displays it with the unit
which is used for manual input in the menu of the sensor with correction
function (e.g. TetraCon
®
700 IQ measures the salinity - salinity is quoted
dimensionless).
– The manually entered value for the static correction is set to the mean
value of the influence quantity (e.g. salinity value) of the test sample.
If no measured value for the influence quantity is available, the correction
is made with the value that was manually entered (static correction).
Establishing the
link
1
Using
<S>
, open the
Settings
menu.
2
Using
<
><
>
and
<OK>
, select and confirm the menu item,
System
settings
->
Sensor-sensor link
.
The
Sensor-sensor link
overview opens.
The displayed list includes all sensors with which the correction of an
influence quantity is possible (e.g. FDO
®
700 IQ).
3
Using
<
><
>
, highlight a sensor and confirm with
<OK>
.
The influence quantity to be linked and the current state of the link are
displayed.