Outputs
DIQ/S 28X
156
ba77168d02
03/2017
7.9
Behavior of linked outputs
7.9.1
Behavior in case of error
For linked relay outputs or current outputs, you can determine the behavior of
the outputs in case of errors.
Depending on the use of the output, the behavior in case of errors is set in the
following menus:
Error events
The specified behavior occurs with the following events or conditions:
The linked sensor does not provide any valid measured value
(display of
Init
,
Error
,
"-----"
, or
OFL)
Communication with the DIQ/S 28X is interrupted for more than 2 minutes.
The supply voltage for the DIQ/S 28X is too low.
In the function
Recorder
, the measured value of the linked sensor is outside
the range between
Start value
and
End value
.
Freezing the
conditions of the
outputs
Irrespective of the behavior in case of error that was defined, the following situ-
ations cause the conditions of the outputs to be frozen:
The linked sensor is in the maintenance condition
(display of
Cal
,
Clean
, or flashing measured value).
Communication with the DIQ/S 28X is temporarily interrupted. After an inter-
ruption of 2 minutes the output changes to the behavior defined for error.
Restoring the
normal functioning
The relay or current output automatically returns to its normal condition as soon
as all errors have been eliminated and all conditions that caused the freezing of
the outputs were eliminated.
Output
Menu
Frequency controller
Frequency with error
Pulse-width output
Error pulse width
(see section 7.6.6)
Recorder
Current with error