Fig. 7-16 Settings of outputs and links
7.7.1
Recorder
Function
The measured values of the linked sensor at the current output are set up as cur-
rent intensity in the
Recorder
application. The output of measured values is laid
down in the
Recorder type
,
Start value
and
End value
settings.
5
Using
<
>
, highlight a line for a current output (Cx) in the column
Feature
. Confirm with
<OK>
.
The
Settings of outputs and links
display opens.
6
Using
<
>
, highlight the menu item,
Current output
, and confirm
with
<OK>
.
7
Select and confirm a function with
<
>
and
<OK>
.
Function
Settings
No function
The current output is not used.
Recorder
PID controller
Fixed current value
8
Make the settings for the current output with
<
>
and
<OK>
.
9
Using
<
>
and
<OK>
, highlight and confirm
Save and quit
.
The new settings are stored.